Long-term retention of cardiac resynchronization therapy
- PMID: 15234410
- DOI: 10.1016/j.jacc.2004.03.054
Long-term retention of cardiac resynchronization therapy
Abstract
Objectives: The purpose of this study was to determine the frequency and causes of intermittent and permanent loss of cardiac resynchronization therapy (CRT) in patients who have undergone the successful implantation of a transvenous defibrillator that delivers CRT (CRT-D).
Background: The causes of loss of CRT have not been described.
Methods: The records of 512 patients who underwent an attempt at implantation of a transvenous CRT-D device as part of the VENTAK CHF/CONTAK CD Biventricular Pacing study were analyzed.
Results: Device implantation was successful in 443 of 512 (87%) of patients. Among these 443 patients, CRT was interrupted in 161 (36%) patients during a mean follow-up of 2.5 +/- 1.1 years. Reasons included the development of an atrial tachyarrhythmia (18%), loss of left ventricular capture (10%), diaphragmatic stimulation (2%), loss of right ventricular capture (2%), infection (1%), intentional discontinuation of CRT (1%), loss of right atrial sensing (1%), and ventricular oversensing (0.2%). Most patients underwent an intervention that permitted the reinstitution of CRT, such that only 20 of the 443 patients (5%) experienced the permanent loss of CRT. Using an intention-to-treat analysis, the long-term retention of CRT was 83% during the course of 2.5 years.
Conclusions: Cardiac resynchronization therapy is interrupted in more than one-third of patients after the successful implantation of a CRT-D device. However, CRT can be reinstituted in most patients and has a high long-term retention rate. Because patients with slower heart rates were more likely to develop atrial tachyarrhythmias, a dual-chamber rate-modulated pacing mode (DDDR) may reduce interruptions of CRT.
Similar articles
-
Predictors of appropriate defibrillator therapy among patients with an implantable defibrillator that delivers cardiac resynchronization therapy.J Cardiovasc Electrophysiol. 2006 May;17(5):486-90. doi: 10.1111/j.1540-8167.2006.00355.x. J Cardiovasc Electrophysiol. 2006. PMID: 16684019 Clinical Trial.
-
The Dual Chamber and VVI Implantable Defibrillator (DAVID) Trial: rationale, design, results, clinical implications and lessons for future trials.Card Electrophysiol Rev. 2003 Dec;7(4):468-72. doi: 10.1023/B:CEPR.0000023165.20987.b1. Card Electrophysiol Rev. 2003. PMID: 15071277 Clinical Trial.
-
Reduced atrial tachyarrhythmia susceptibility after upgrade of conventional implanted pulse generator to cardiac resynchronization therapy in patients with heart failure.J Am Coll Cardiol. 2007 Sep 25;50(13):1246-51. doi: 10.1016/j.jacc.2007.04.096. Epub 2007 Sep 10. J Am Coll Cardiol. 2007. PMID: 17888841
-
Cardiac resynchronization therapy: device-based medicine for heart failure.J Card Surg. 2004 May-Jun;19(3):270-4. doi: 10.1111/j.0886-0440.2004.04081.x. J Card Surg. 2004. PMID: 15151661 Review.
-
Advances in cardiac pacing: beyond the transvenous right ventricular apical lead.Cardiovasc Ther. 2010 Dec;28(6):369-79. doi: 10.1111/j.1755-5922.2010.00157.x. Cardiovasc Ther. 2010. PMID: 20553288 Review.
Cited by
-
Discontinuation of Cardiac Resynchronization Therapy for Heart Failure Due to Dilated Cardiomyopathy in a 61-Year-Old Female "-Super-Responder" with Return of a Reduced Left Ventricular Ejection Fraction to Normal.Am J Case Rep. 2020 Sep 28;21:e926704. doi: 10.12659/AJCR.926704. Am J Case Rep. 2020. PMID: 32986689 Free PMC article.
-
Phrenic nerve stimulation in cardiac resynchronization therapy.J Interv Card Electrophysiol. 2014 Oct;41(1):15-21. doi: 10.1007/s10840-014-9917-8. Epub 2014 Jun 17. J Interv Card Electrophysiol. 2014. PMID: 24934757 Review.
-
Rate Control in Atrial Fibrillation: Methods for Assessment, Targets for Ventricular Rate during AF, and Clinical Relevance for Device Therapy.J Atr Fibrillation. 2013 Jun 30;6(1):791. doi: 10.4022/jafib.791. eCollection 2013 Jun-Jul. J Atr Fibrillation. 2013. PMID: 28496852 Free PMC article. Review.
-
Performance and clinical comparison between left ventricular quadripolar and bipolar leads in cardiac resynchronization therapy: Observational research.Indian Heart J. 2018 Nov-Dec;70(6):864-871. doi: 10.1016/j.ihj.2018.05.007. Epub 2018 May 16. Indian Heart J. 2018. PMID: 30580858 Free PMC article.
-
Comparison of different pacing strategies to minimize phrenic nerve stimulation in cardiac resynchronization therapy.J Cardiovasc Electrophysiol. 2013 Sep;24(9):1008-14. doi: 10.1111/jce.12159. Epub 2013 Apr 26. J Cardiovasc Electrophysiol. 2013. PMID: 23621543 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical
Research Materials