Laboratory evaluation of heat-and-moisture exchangers
- PMID: 15236061
- DOI: 10.1007/s0054090030080
Laboratory evaluation of heat-and-moisture exchangers
Abstract
We conducted a laboratory study on six commercially available heat and moisture exchangers in order to determine and compare their water retaining efficiency and their contribution to airway resistance. The Gambro-Engström Edith Flex device was the most desirable of the six devices we evaluated in terms of its water retaining efficiency. The NMI Pneumoist 1 and the Siemens Servo Humidifier 153 units had good water retaining capacity but their higher airflow resistance need close monitoring, especially after prolonged clinical use. The Pall HME 15-22 and the Portex Humid-Vent 1 devices were also efficient in water retaining capacity. The Pall also demonstrated low airflow resistance and the minimum increase in airflow resistance after water immersion. The pathogen filtering capacity of the Pall should also be considered an additional advantage, especially in infected patients. The Terumo Breathaid device performed worst of all six devices, but it was still better than no HME at all.
Similar articles
-
Efficacy of nine heat and moisture exchangers for intraoperative airway heat conservation.J Aerosol Med. 1991 Summer;4(2):117-25. doi: 10.1089/jam.1991.4.117. J Aerosol Med. 1991. PMID: 10147688
-
Performance evaluation of six heat and moisture exchangers according to the Draft International Standard (ISO/DIS 9360).Acta Anaesthesiol Scand. 1990 Jul;34(5):404-9. doi: 10.1111/j.1399-6576.1990.tb03112.x. Acta Anaesthesiol Scand. 1990. PMID: 2389657
-
Preservation of humidity and heat of respiratory gases in patients with a minute ventilation greater than 10 L/min.Crit Care Med. 1994 Nov;22(11):1871-6. Crit Care Med. 1994. PMID: 7956294 Clinical Trial.
-
An evaluation of six disposable heat and moisture exchangers.Anaesth Intensive Care. 1987 Aug;15(3):317-22. doi: 10.1177/0310057X8701500312. Anaesth Intensive Care. 1987. PMID: 3661965
-
Heat and moisture exchangers and breathing system filters: their use in anaesthesia and intensive care. Part 2 - practical use, including problems, and their use with paediatric patients.Anaesthesia. 2011 Jan;66(1):40-51. doi: 10.1111/j.1365-2044.2010.06564.x. Epub 2010 Nov 30. Anaesthesia. 2011. PMID: 21118189 Review.
Cited by
-
Ventilator-associated pneumonia using a heated humidifier or a heat and moisture exchanger: a randomized controlled trial [ISRCTN88724583].Crit Care. 2006;10(4):R116. doi: 10.1186/cc5009. Crit Care. 2006. PMID: 16884530 Free PMC article. Clinical Trial.
-
A novel method of evaluation of three heat-moisture exchangers in six different ventilator settings.Intensive Care Med. 1998 Feb;24(2):138-46. doi: 10.1007/s001340050535. Intensive Care Med. 1998. PMID: 9539071