Is galvanic corrosion between titanium alloy and stainless steel spinal implants a clinical concern?
- PMID: 15246296
- DOI: 10.1016/j.spinee.2003.12.004
Is galvanic corrosion between titanium alloy and stainless steel spinal implants a clinical concern?
Abstract
Background context: Surgeons are hesitant to mix components made of differing metal classes for fear of galvanic corrosion complications. However, in vitro studies have failed to show a significant potential for galvanic corrosion between titanium and stainless steel, the two primary metallic alloys used for spinal implants. Galvanic corrosion resulting from metal mixing has not been described in the literature for spinal implant systems.
Purpose: To determine whether galvanic potential significantly affects in vitro corrosion of titanium and stainless steel spinal implant components during cyclical compression bending.
Study design/setting: Bilateral spinal implant constructs consisting of pedicle screws, slotted connectors, 6.35-mm diameter rods and a transverse rod connector assembled in polyethylene test blocks were tested in vitro. Two constructs had stainless steel rods with mixed stainless steel (SS-SS) and titanium (SS-Ti) components, and two constructs had titanium rods with mixed stainless steel (Ti-SS) and titanium (Ti-Ti) components.
Methods: Each construct was immersed in phosphate-buffered saline (pH 7.4) at 37 C and tested in cyclic compression bending using a sinusoidal load-controlling function with a peak load of 300 N and a frequency of 5 Hz until a level of 5 million cycles was reached. The samples were then removed and analyzed visually for evidence of corrosion. In addition, scanning electron microscopy (SEM) and energy dispersive spectrometry (EDS) were used to evaluate the extent of corrosion at the interconnections.
Results: None of the constructs failed during testing. Gross observation of the implant components after disassembly revealed that no corrosion had occurred on the surface of the implants that had not been in contact with another component. The Ti-Ti interfaces showed some minor signs of corrosion only detectable using SEM and EDS. The greatest amount of corrosion occurred at the SS-SS interfaces and was qualitatively less at the SS-Ti and Ti-SS interfaces.
Conclusions: The results from this study indicate that when loaded dynamically in saline, stainless steel implant components have a greater susceptibility to corrosion than titanium. Furthermore, the galvanic potential between the dissimilar metals does not cause a discernible effect on the corrosion of either. Although the mixture of titanium alloy with stainless steel is not advocated, the results of this study suggest that galvanic corrosion is less pronounced in SS-Ti mixed interfaces than in all stainless steel constructs.
Similar articles
-
Comparison between 4.0-mm stainless steel and 4.75-mm titanium alloy single-rod spinal instrumentation for anterior thoracoscopic scoliosis surgery.Spine (Phila Pa 1976). 2008 Sep 15;33(20):2173-8. doi: 10.1097/BRS.0b013e31817f9415. Spine (Phila Pa 1976). 2008. PMID: 18794758
-
The effects of rod contouring on spinal construct fatigue strength.Spine (Phila Pa 1976). 2006 Jul 1;31(15):1680-7. doi: 10.1097/01.brs.0000224177.97846.00. Spine (Phila Pa 1976). 2006. PMID: 16816763
-
Should the galvanic combination of titanium and stainless steel surgical implants be avoided?Injury. 2008 Feb;39(2):161-9. doi: 10.1016/j.injury.2007.07.015. Epub 2007 Dec 3. Injury. 2008. PMID: 18054018
-
[Biodeterioration and corrosion of metallic implants and prostheses].Medicina (B Aires). 1993;53(3):260-74. Medicina (B Aires). 1993. PMID: 8114635 Review. Spanish.
-
An Overview of Internal Fixation Implant Metallurgy and Galvanic Corrosion Effects.J Hand Surg Am. 2015 Aug;40(8):1703-10, 1710.e1-4. doi: 10.1016/j.jhsa.2015.03.030. Epub 2015 Jul 1. J Hand Surg Am. 2015. PMID: 26143030 Review.
Cited by
-
Serum metal ion levels following spinal deformity surgery: a case-control study of 182 individuals.Eur Spine J. 2022 Nov;31(11):3036-3041. doi: 10.1007/s00586-022-07341-5. Epub 2022 Sep 14. Eur Spine J. 2022. PMID: 36102991
-
Total femoral spanning for distal femur "fragility" fractures utilising nail-plate fixation "short-term experience of a district general hospital".Eur J Orthop Surg Traumatol. 2024 May;34(4):2003-2013. doi: 10.1007/s00590-024-03883-0. Epub 2024 Mar 20. Eur J Orthop Surg Traumatol. 2024. PMID: 38509381
-
Integration of heterologous bone matrix associated with polymethylmethacrylate in induced tibial bone defects. An experimental study in rabbits.Acta Cir Bras. 2021 Sep 3;36(7):e360704. doi: 10.1590/ACB360704. eCollection 2021. Acta Cir Bras. 2021. PMID: 34495139 Free PMC article.
-
Indications for implant removal after fracture healing: a review of the literature.Eur J Trauma Emerg Surg. 2013 Aug;39(4):327-37. doi: 10.1007/s00068-013-0283-5. Epub 2013 Apr 12. Eur J Trauma Emerg Surg. 2013. PMID: 26815392
-
Pseudarthrosis due to galvanic corrosion presenting as subarachnoid hemorrhage.J Craniovertebr Junction Spine. 2017 Apr-Jun;8(2):156-158. doi: 10.4103/jcvjs.JCVJS_6_17. J Craniovertebr Junction Spine. 2017. PMID: 28694602 Free PMC article.