Source configuration and dose rates for the Selectron afterloading equipment for gynecologic applicators
- PMID: 1526871
- DOI: 10.1016/0360-3016(92)90688-e
Source configuration and dose rates for the Selectron afterloading equipment for gynecologic applicators
Abstract
This paper describes source train configurations of the Selectron LDR-6 that can be used to duplicate commonly used manually afterloadable source configurations for gynecological intracavitary brachytherapy. We demonstrate that the recommended source configuration using 2.5 mm 137Cs spheres is dosimetrically equivalent to its manually afterloaded counterparts using 3M 137Cs sources (20 mm overall length, 14 mm active length). Isodose distributions are in agreement to within 5% for the Fletcher-Suit-Delclos family of tandems, colpostats, and vaginal cylinders. Our data also demonstrate that the isotropic point source model used by most computerized treatment planning systems is a good approximation to source trains of filtered, spherical sources interspersed with solid steel spacers. We conclude that the use of 2.5 mm 137Cs sources in the Selectron remote afterloading system can accurately reproduce the isodose distribution achieved with linear 137Cs manually afterloaded sources.
Similar articles
-
Gynaecological treatments using the Selectron remote afterloading system.Br J Radiol. 1988 Nov;61(731):1053-7. doi: 10.1259/0007-1285-61-731-1053. Br J Radiol. 1988. PMID: 3208009
-
EGSnrc-based Monte Carlo dosimetry of CSA1 and CSA2 137Cs brachytherapy source models.Med Phys. 2009 Sep;36(9):3870-9. doi: 10.1118/1.3176936. Med Phys. 2009. PMID: 19810459
-
AAPM TG-43 formalism for brachytherapy dose calculation of a 137Cs tube source.Med Phys. 2004 Apr;31(4):755-9. doi: 10.1118/1.1650564. Med Phys. 2004. PMID: 15124992
-
[Pulsed dose brachytherapy--a review of clinical applications in the treatment of gynecological malignancies].Ginekol Pol. 2005 Aug;76(8):661-70. Ginekol Pol. 2005. PMID: 16363375 Review. Polish.
-
[Review on the afterloading technics in gynecological radiotherapy].Strahlentherapie. 1975 Sep;150(3):237-42. Strahlentherapie. 1975. PMID: 1108304 Review. German.
Cited by
-
Impact of the vaginal applicator and dummy pellets on the dosimetry parameters of Cs-137 brachytherapy source.J Appl Clin Med Phys. 2011 May 19;12(3):3480. doi: 10.1120/jacmp.v12i3.3480. J Appl Clin Med Phys. 2011. PMID: 21844861 Free PMC article.
-
Dose distribution verification for GYN brachytherapy using EBT Gafchromic film and TG-43 calculation.Rep Pract Oncol Radiother. 2016 Sep-Oct;21(5):480-6. doi: 10.1016/j.rpor.2016.06.005. Epub 2016 Jul 18. Rep Pract Oncol Radiother. 2016. PMID: 27489519 Free PMC article.
-
Independent evaluation of an in-house brachytherapy treatment planning system using simulation, measurement and calculation methods.J Appl Clin Med Phys. 2012 Mar 8;13(2):3687. doi: 10.1120/jacmp.v13i2.3687. J Appl Clin Med Phys. 2012. PMID: 22402384 Free PMC article.
-
Perturbation of TG-43 parameters of the brachytherapy sources under insufficient scattering materials.J Appl Clin Med Phys. 2013 May 6;14(3):4228. doi: 10.1120/jacmp.v14i3.4228. J Appl Clin Med Phys. 2013. PMID: 23652255 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources