Single-shot half-Fourier RARE sequence with ultra-short inter-echo spacing for lung imaging
- PMID: 15269963
- DOI: 10.1002/jmri.20107
Single-shot half-Fourier RARE sequence with ultra-short inter-echo spacing for lung imaging
Abstract
Purpose: To improve the image quality of pulmonary magnetic resonance (MR) imaging using an ultra-short inter-echo spacing half-Fourier single shot rapid acquisition with relaxation enhancement (USHA-RARE) sequence.
Materials and methods: Pulmonary MR images were acquired by USHA-RARE sequence with various inter-echo spacings. The sequence parameters were as follows: repetition time (TR)/effective TE: infinite/39-41 msec; section thickness: 10 mm; acquisition matrix: 128 x 128; field of view: 450 x 450 mm. Inter-echo spacing varied (2.5 msec, 3.0 msec, 3.5 msec, 4.0 msec, 4.5 msec, 5.0 msec), and the respective phase-encoding steps were 80, 77, 75, 74, 73, and 72. Signal-to-noise ratios (SNRs), the signal ratios between lung and fat (lung-to-fat ratio: LFRs), and the signal ratios between the lung and the serratus anterior muscle (lung-to-muscle ratio: LMRs) of each inter-echo spacing were calculated, and statistically evaluated.
Results: The SNRs at inter-echo spacings of < or = 3.0 msec were significantly higher than those > or = 4.0 msec (P < 0.05). The LFRs and LMRs at inter-echo spacing < or = 3.0 msec were significantly higher than those > or = 4.0 msec (P < 0.05).
Conclusion: USHA-RARE sequence does improve signal intensity from the lung.
Copyright 2004 Wiley-Liss, Inc.
Similar articles
-
Centrically reordered inversion recovery half-Fourier single-shot turbo spin-echo sequence: improvement of the image quality of oxygen-enhanced MRI.Eur J Radiol. 2004 Nov;52(2):200-5. doi: 10.1016/j.ejrad.2004.02.007. Eur J Radiol. 2004. PMID: 15489080
-
Influence of inversion pulse type in assessing lung-oxygen-enhancement by centrically-reordered non-slice-selective inversion-recovery half-Fourier single-shot turbo spin-echo (HASTE) sequence.J Magn Reson Imaging. 2007 Oct;26(4):1133-8. doi: 10.1002/jmri.21105. J Magn Reson Imaging. 2007. PMID: 17896357
-
MR imaging of the lungs: value of short TE spin-echo pulse sequences.AJR Am J Roentgenol. 1992 Nov;159(5):951-6. doi: 10.2214/ajr.159.5.1414805. AJR Am J Roentgenol. 1992. PMID: 1414805
-
Quantitative regional oxygen transfer imaging of the human lung.J Magn Reson Imaging. 2007 Sep;26(3):637-45. doi: 10.1002/jmri.21033. J Magn Reson Imaging. 2007. PMID: 17685416
-
Phase shift due to inversion recovery process and chemical shift effect in inversion recovery prepared gradient-echo magnetic resonance imaging.Radiat Med. 2005 Feb;23(1):56-60. Radiat Med. 2005. PMID: 15786753
Cited by
-
Lung Parenchymal Signal Intensity in MRI: A Technical Review with Educational Aspirations Regarding Reversible Versus Irreversible Transverse Relaxation Effects in Common Pulse Sequences.Concepts Magn Reson Part A Bridg Educ Res. 2014 Mar 1;43A(2):29-53. doi: 10.1002/cmr.a.21297. Concepts Magn Reson Part A Bridg Educ Res. 2014. PMID: 25228852 Free PMC article.
-
Free-Breathing Phase-Resolved Oxygen-Enhanced Pulmonary MRI Based on 3D Stack-of-Stars UTE Sequence.Sensors (Basel). 2022 Apr 24;22(9):3270. doi: 10.3390/s22093270. Sensors (Basel). 2022. PMID: 35590959 Free PMC article.
-
Gas Phase UTE MRI of Propane and Propene.Tomography. 2016 Mar;2(1):49-55. doi: 10.18383/j.tom.2016.00112. Tomography. 2016. PMID: 27478870 Free PMC article.
-
Clinical PET-MR Imaging in Breast Cancer and Lung Cancer.PET Clin. 2016 Oct;11(4):387-402. doi: 10.1016/j.cpet.2016.05.008. PET Clin. 2016. PMID: 27593245 Free PMC article. Review.
-
State-of-the-art MR Imaging for Thoracic Diseases.Magn Reson Med Sci. 2022 Mar 1;21(1):212-234. doi: 10.2463/mrms.rev.2020-0184. Epub 2021 Apr 29. Magn Reson Med Sci. 2022. PMID: 33952785 Free PMC article. Review.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical