An evaluation of the heat and moisture exchange performance of four ventilator circuit filters
- PMID: 1527255
- DOI: 10.1007/BF01706470
An evaluation of the heat and moisture exchange performance of four ventilator circuit filters
Erratum in
- Intensive Care Med 1992;18(8):498
Abstract
Objective: To compare the heat and moisture exchange efficiency of 4 commonly used ventilator circuit filters--DAR Hygroster (DHS), DAR Hygobac (DHC), Pall Ultipor BB50 (PUBC) and Intersurgical filtatherm (IFT).
Design: Prospective randomized study.
Setting: Intensive care unit of a university teaching hospital.
Patients and participants: 80 patients requiring post-operative mechanical ventilation were entered into the study. A total of 40 patients were studied after 1 h and 40 after 24 h. Due to technical errors 6 of the 1 h studies and 2 of the 24 h studies were excluded from analysis.
Interventions: Mean wet and dry ceramic-platinum thermal probes were inserted in the inspiratory side of a device which separated inspiratory and expiratory gas flows on the patient side of the ventilator circuit filter. Humidification efficiency was calculated with reference to these wet and dry bulb temperatures. Tracheal temperature was measured by a similar thermal probe in the tracheal tube.
Measurements and results: Absolute humidity was greatest in group DHS (30.4 +/- 3.5 versus 22.9 +/- 1.5 mg/l at 1 h and 27.3 +/- 3.5 versus 20.7 +/- 2.4 mg/l at 24 h). The differences in absolute humidity reached statistical significance as follows: p = 0.0001 for DHS versus all others at 1 h and for DHS and DHC versus PUBC and IFT at 24 h.
Conclusion: Better humidification capabilities were obtained with groups DHS and DHC than PUBC and IFT.
Similar articles
-
Comparing two heat and moisture exchangers with one vaporizing humidifier in patients with minute ventilation greater than 10 L/min.Chest. 1995 May;107(5):1411-5. doi: 10.1378/chest.107.5.1411. Chest. 1995. PMID: 7750340 Clinical Trial.
-
Preservation of humidity and heat of respiratory gases in patients with a minute ventilation greater than 10 L/min.Crit Care Med. 1994 Nov;22(11):1871-6. Crit Care Med. 1994. PMID: 7956294 Clinical Trial.
-
A novel method of evaluation of three heat-moisture exchangers in six different ventilator settings.Intensive Care Med. 1998 Feb;24(2):138-46. doi: 10.1007/s001340050535. Intensive Care Med. 1998. PMID: 9539071
-
Performance evaluation of three vaporizing humidifiers and two heat and moisture exchangers in patients with minute ventilation > 10 L/min.Chest. 1992 Nov;102(5):1347-50. doi: 10.1378/chest.102.5.1347. Chest. 1992. PMID: 1424849 Clinical Trial.
-
The effects of excessive humidity.Respir Care Clin N Am. 1998 Jun;4(2):215-28. Respir Care Clin N Am. 1998. PMID: 9648183 Review.
Cited by
-
Ventilator-associated pneumonia using a heated humidifier or a heat and moisture exchanger: a randomized controlled trial [ISRCTN88724583].Crit Care. 2006;10(4):R116. doi: 10.1186/cc5009. Crit Care. 2006. PMID: 16884530 Free PMC article. Clinical Trial.
-
Factors associated with blocked tracheal tubes.Intensive Care Med. 2001 Oct;27(10):1679-81. doi: 10.1007/s001340101079. Intensive Care Med. 2001. PMID: 11685314 No abstract available.
-
An evaluation of the microbial retention performance of three ventilator-circuit filters.Intensive Care Med. 1994;20(3):233-7. doi: 10.1007/BF01704708. Intensive Care Med. 1994. PMID: 8014294 Clinical Trial.
-
Evaluating humidity recovery efficiency of currently available heat and moisture exchangers: a respiratory system model study.Clinics (Sao Paulo). 2009;64(6):585-90. doi: 10.1590/s1807-59322009000600015. Clinics (Sao Paulo). 2009. PMID: 19578664 Free PMC article.
References
Publication types
MeSH terms
LinkOut - more resources
Other Literature Sources