Spontaneous minute ventilation is a predictor of extubation failure in extremely-low-birth-weight infants
- PMID: 15280139
- DOI: 10.1080/14767050410001668239
Spontaneous minute ventilation is a predictor of extubation failure in extremely-low-birth-weight infants
Abstract
Objective: To validate the percentage of time spent below a target value of spontaneous expiratory minute ventilation (< 125 ml/min per kg) during a 2-h period of continuous positive airway pressure (CPAP) via an endotracheal tube (ETT) as a predictor of failed extubation in preterm infants.
Methods: Forty-one infants intubated for at least 24 h, with birth weight between 500 and 1000 g, who were clinically stable and at ventilator setting compatible with an extubation attempt, were studied during a 2-h period of ETT CPAP. Dynamic lung compliance and total lung resistance were measured during a period of quiet breathing, while tidal volume (Vt), respiratory rate and the corresponding spontaneous expiratory minute ventilation values were calculated for the complete recording period of 2 h using a customized computer program. The time each patient spent below the target spontaneous expiratory minute ventilation value was reported as a percentage of the total recorded time (% spontaneous expiratory minute ventilation < 125 ml/min per kg). Extubation failure was defined as the need for reintubation within 72 h.
Results: Eleven out of 41 babies (26.8%) experienced failure of extubation (failure group) while 30 infants (73.2%) were successfully extubated (success group). There were no significant differences in dynamic lung compliance and lung resistance between the two groups, but the mean values of respiratory rate and spontaneous expiratory minute ventilation were significantly lower in the failure group than in the success group: 43 (37-56) breaths/min and 240 (160-353) ml/min per kg vs. 53 (28-67) breaths/min and 309 (223-434) ml/min per kg, respectively (p = 0.0129 and p = 0.0039). Moreover, the babies in whom extubation failed spent a longer time below the target value of spontaneous expiratory minute ventilation when compared with successfully extubated babies (p < 0.0001). Percentage of time spent with spontaneous expiratory minute ventilation < 125 ml/min per kg had a larger area than transcutaneous (Tc)PCO2, TcPO2 and pulse oxymetry saturation (SpO2) under the receiver operator characteristic curves.
Conclusion: The measurement of spontaneous expiratory minute ventilation prior to extubation could be useful in identifying those babies who are not ready for spontaneous ventilation.
Similar articles
-
Compliance of the respiratory system as a predictor for successful extubation in very-low-birth-weight infants recovering from respiratory distress syndrome.S Afr Med J. 1999 Oct;89(10):1097-102. S Afr Med J. 1999. PMID: 10582068
-
Predictors of successful extubation of preterm low-birth-weight infants with respiratory distress syndrome.Pediatr Crit Care Med. 2005 Jan;6(1):44-9. doi: 10.1097/01.PCC.0000149136.28598.14. Pediatr Crit Care Med. 2005. PMID: 15636658
-
A decision-tree model for predicting extubation outcome in elderly patients after a successful spontaneous breathing trial.Anesth Analg. 2010 Nov;111(5):1211-8. doi: 10.1213/ANE.0b013e3181f4e82e. Epub 2010 Sep 14. Anesth Analg. 2010. PMID: 20841406
-
[Pulmonary function in preterm and full term infants during the neonatal period: 1. respiratory pattern].An Esp Pediatr. 1991 Oct;35(4):243-7. An Esp Pediatr. 1991. PMID: 1763850 Review. Spanish.
-
[Pulmonary function in the full-term newborn infant without respiratory problems].An Esp Pediatr. 1990 Dec;33(6):511-4. An Esp Pediatr. 1990. PMID: 2077989 Review. Spanish.
Cited by
-
Respiratory Management of the Preterm Infant: Supporting Evidence-Based Practice at the Bedside.Children (Basel). 2023 Mar 10;10(3):535. doi: 10.3390/children10030535. Children (Basel). 2023. PMID: 36980093 Free PMC article. Review.
-
Prediction of Extubation readiness in extremely preterm infants by the automated analysis of cardiorespiratory behavior: study protocol.BMC Pediatr. 2017 Jul 17;17(1):167. doi: 10.1186/s12887-017-0911-z. BMC Pediatr. 2017. PMID: 28716018 Free PMC article.
-
Predictors of successful extubation from volume-targeted ventilation in extremely preterm neonates.J Perinatol. 2024 Feb;44(2):250-256. doi: 10.1038/s41372-023-01849-4. Epub 2023 Dec 20. J Perinatol. 2024. PMID: 38123799
-
Extubation from high-frequency oscillatory ventilation in extremely low birth weight infants: a prospective observational study.BMJ Paediatr Open. 2018 Nov 9;2(1):e000350. doi: 10.1136/bmjpo-2018-000350. eCollection 2018. BMJ Paediatr Open. 2018. PMID: 30498796 Free PMC article.
-
Diaphragmatic electromyography during a spontaneous breathing trial to predict extubation failure in preterm infants.Pediatr Res. 2022 Oct;92(4):1064-1069. doi: 10.1038/s41390-022-02085-w. Epub 2022 May 6. Pediatr Res. 2022. PMID: 35523885 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources