Polyethylene post failure in posterior stabilized total knee arthroplasty
- PMID: 15284990
- DOI: 10.1016/j.arth.2004.02.026
Polyethylene post failure in posterior stabilized total knee arthroplasty
Abstract
Posterior-stabilized (PS) prostheses have been used extensively in total knee arthroplasty (TKA), with excellent long-term results. The key feature of these prostheses is the femoral cam and tibial post mechanism that limits posterior displacement and produces femoral rollback. Although articular-surface polyethylene wear of the tibial component has not been a significant clinical problem, tibial post wear has been reported. In distinction to chronic post wear, little information exists about catastrophic post failure. We present the case of a 56-year-old woman who presented 63 months after TKA with a PS prosthesis with acute fracture of the polyethylene post. The evaluation and treatment of this patient, including the previously unreported use of computed tomography arthrography to diagnose this rare problem, is reviewed.
Similar articles
-
Fracture of the polyethylene tibial post in the posterior-stabilized total knee prosthesis: arthrographic and CT arthrographic diagnosis.Skeletal Radiol. 2019 Jul;48(7):1145-1148. doi: 10.1007/s00256-019-03173-5. Epub 2019 Feb 7. Skeletal Radiol. 2019. PMID: 30729263
-
Fractured polyethylene tibial post in a posterior-stabilized knee prosthesis presenting as a floating palpable mass.J Knee Surg. 2009 Oct;22(4):374-6. doi: 10.1055/s-0030-1247780. J Knee Surg. 2009. PMID: 19902738
-
Fracture of a polyethylene tibial post in a Scorpio posterior-stabilized knee prosthesis.Clin Orthop Surg. 2009 Jun;1(2):118-21. doi: 10.4055/cios.2009.1.2.118. Epub 2009 May 27. Clin Orthop Surg. 2009. PMID: 19885065 Free PMC article.
-
The use of highly cross-linked polyethylene in total knee arthroplasty.J Am Acad Orthop Surg. 2011 Mar;19(3):143-51. doi: 10.5435/00124635-201103000-00003. J Am Acad Orthop Surg. 2011. PMID: 21368095 Review.
-
Cemented posterior stabilized total knee arthroplasty.J Arthroplasty. 2004 Jun;19(4 Suppl 1):17-21. doi: 10.1016/j.arth.2004.02.014. J Arthroplasty. 2004. PMID: 15190543 Review.
Cited by
-
How to treat a tibial post fracture in total knee arthroplasty? A systematic review.Clin Orthop Relat Res. 2011 Jun;469(6):1709-15. doi: 10.1007/s11999-010-1609-9. Epub 2010 Oct 21. Clin Orthop Relat Res. 2011. PMID: 20963534 Free PMC article.
-
Patellar Dislocation Masking a Tibial Post Fracture: An Unusual Complication of Posterior Stabilized Total Knee Arthroplasty.Cureus. 2023 Nov 17;15(11):e48950. doi: 10.7759/cureus.48950. eCollection 2023 Nov. Cureus. 2023. PMID: 38106799 Free PMC article.
-
Erosion and Failure of the Tibial Post after Posterior-stabilized Total Knee Replacement: A Case Report.J Orthop Case Rep. 2023 Oct;13(10):127-131. doi: 10.13107/jocr.2023.v13.i10.3962. J Orthop Case Rep. 2023. PMID: 37885652 Free PMC article.
-
Nontraumatic fracture of the femoral condylar prosthesis in a total knee arthroplasty leading to mechanical failure.Case Rep Orthop. 2014;2014:896348. doi: 10.1155/2014/896348. Epub 2014 Jan 23. Case Rep Orthop. 2014. PMID: 24587928 Free PMC article.
-