Susceptibility of automated external defibrillators to train overhead lines and metro third rails
- PMID: 15294405
- DOI: 10.1016/j.resuscitation.2004.02.018
Susceptibility of automated external defibrillators to train overhead lines and metro third rails
Abstract
Introduction: Immediate accessibility to automated external defibrillators (AED) is recommended for highly frequented public areas. In train terminals and metro stations electromagnetic interference (EMI) is present. In preparation for a public access defibrillation (PAD) programme in this environment possible effects on AED safety and accuracy were studied.
Methods: In typical public transportation settings 11 different AED models were bench tested for their sensitivity and specificity of ECG analysis with shockable and nonshockable rhythms provided by an ECG simulator. The devices were exposed to the electromagnetic interference of a rail system operating with 15 kV alternating current (ac) with a frequency of 16 2/3 Hz and a subway system powered with 750 V direct current (dc). AED cables were setup parallel and perpendicular to the tracks, the tests were carried out at 3 m distance from the rails in an empty station and with incoming trains.
Results: A total of 5280 tests were recorded, each device was tested a total of 480 times. Fifteen kilovolts 16 2/3 Hz ac interfered more than 750 V dc with the tachyarrhythmia detection systems (P < 0.0001). An AED setup with electrode cables perpendicular to track and power line reduced interference (P < 0.0001), while incoming trains had no significant effect on ECG analysis (P = 0.19). Depending on the AED model, sensitivity ranged from 60 to 100% and specificity from 54 to 100%, representing a positive likelihood-ratio from 1.3 to 241 and a negative likelihood-ratio from 0.7 to 0.0. In the public transportation setting tested, four AED models were unsuitable for automated defibrillation as these devices demonstrated an unacceptable performance in respect of accuracy and safety. In the train setting two devices performed with an accuracy of 57 and 65%. One AED recommended shocks for sinus rhythm at normal frequency. In the metro setting one AED did not advise shocks for ventricular tachycardia.
Conclusion: Shock advisory systems of some AED models are susceptible to electromagnetic interference, especially in terminals with 15 kV 16 2/3 Hz ac power supplies. Interference is minimized, if patient position is parallel and electrode cables are perpendicular to overhead line. The choice of AED model for train or metro stations depends on its lack of susceptibility to typical electromagnetic interference.
Comment in
-
Influence of electromagnetic interference on the human electrocardiogram in automated external defibrillators.Resuscitation. 2005 Jul;66(1):105-6; author reply 106-7. doi: 10.1016/j.resuscitation.2005.01.015. Resuscitation. 2005. PMID: 15993736 No abstract available.
Similar articles
-
Safety aspects for public access defibrillation using automated external defibrillators near high-voltage power lines.Acta Anaesthesiol Scand. 2004 May;48(5):595-600. doi: 10.1111/j.1399-6576.2004.00370.x. Acta Anaesthesiol Scand. 2004. PMID: 15101855
-
Automated external defibrillators do not recommend false positive shocks under the influence of electromagnetic fields present at public locations.Anesth Analg. 2006 Dec;103(6):1485-8. doi: 10.1213/01.ane.0000244477.15072.16. Anesth Analg. 2006. PMID: 17122228 Clinical Trial.
-
Bench study of the accuracy of a commercial AED arrhythmia analysis algorithm in the presence of electromagnetic interferences.Physiol Meas. 2009 Jul;30(7):695-705. doi: 10.1088/0967-3334/30/7/012. Epub 2009 Jun 12. Physiol Meas. 2009. PMID: 19525573
-
[Automatic external defibrillator (AED) and public access defibrillation (PAD)].Harefuah. 2006 May;145(5):375-9, 396. Harefuah. 2006. PMID: 16805222 Review. Hebrew.
-
Implantable rhythm devices and electromagnetic interference: myth or reality?Expert Rev Cardiovasc Ther. 2008 Jul;6(6):823-32. doi: 10.1586/14779072.6.6.823. Expert Rev Cardiovasc Ther. 2008. PMID: 18570620 Review.
Cited by
-
A simple method for removing initial irregularity of an electrocardiogram during a transient state of a power supply in a defibrillator.Technol Health Care. 2020;28(S1):327-334. doi: 10.3233/THC-209033. Technol Health Care. 2020. PMID: 32364165 Free PMC article.
-
Performance of cardiopulmonary resuscitation feedback systems in a long-distance train with distributed traction.Technol Health Care. 2018;26(3):529-535. doi: 10.3233/THC-181241. Technol Health Care. 2018. PMID: 29710761 Free PMC article.
-
[Technical requirements for early defibrillation: what are the capabilities of automated external defibrillators].Herzschrittmacherther Elektrophysiol. 2005 Jun;16(2):84-93. doi: 10.1007/s00399-005-0468-7. Herzschrittmacherther Elektrophysiol. 2005. PMID: 15997355 Review. German.
-
Nociception in Chicken Embryos, Part II: Embryonal Development of Electroencephalic Neuronal Activity In Ovo as a Prerequisite for Nociception.Animals (Basel). 2023 Sep 7;13(18):2839. doi: 10.3390/ani13182839. Animals (Basel). 2023. PMID: 37760239 Free PMC article.
-
Suppression of AC railway power-line interference in ECG signals recorded by public access defibrillators.Biomed Eng Online. 2005 Nov 26;4:65. doi: 10.1186/1475-925X-4-65. Biomed Eng Online. 2005. PMID: 16309558 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources
Medical
Miscellaneous