Combining Ar ion milling with FIB lift-out techniques to prepare high quality site-specific TEM samples
- PMID: 15312185
- DOI: 10.1111/j.0022-2720.2004.01376.x
Combining Ar ion milling with FIB lift-out techniques to prepare high quality site-specific TEM samples
Abstract
Focused ion beam (FIB) techniques can prepare site-specific transmission electron microscopy (TEM) cross-section samples very quickly but they suffer from beam damage by the high energy Ga+ ion beam. An amorphous layer about 20-30 nm thick on each side of the TEM lamella and the supporting carbon film makes FIB-prepared samples inferior to the traditional Ar+ thinned samples for some investigations such as high resolution transmission electron microscopy (HRTEM) and electron energy loss spectroscopy (EELS). We have developed techniques to combine broad argon ion milling with focused ion beam lift-out methods to prepare high-quality site-specific TEM cross-section samples. Site-specific TEM cross-sections were prepared by FIB and lifted out using a Narishige micromanipulator onto a half copper-grid coated with carbon film. Pt deposition by FIB was used to bond the lamellae to the Cu grid, then the coating carbon film was removed and the sample on the bare Cu grid was polished by the usual broad beam Ar+ milling. By doing so, the thickness of the surface amorphous layers is reduced substantially and the sample quality for TEM observation is as good as the traditional Ar+ milled samples.
Similar articles
-
Combining FIB milling and conventional Argon ion milling techniques to prepare high-quality site-specific TEM samples for quantitative EELS analysis of oxygen in molten iron.J Microsc. 2010 Jun 1;238(3):200-9. doi: 10.1111/j.1365-2818.2009.03341.x. J Microsc. 2010. PMID: 20579258
-
Focused ion beam milling and ultramicrotomy of mineralised ivory dentine for analytical transmission electron microscopy.Micron. 2009 Jun;40(4):495-501. doi: 10.1016/j.micron.2008.12.002. Epub 2008 Dec 24. Micron. 2009. PMID: 19157888
-
Applications of the FIB lift-out technique for TEM specimen preparation.Microsc Res Tech. 1998 May 15;41(4):285-90. doi: 10.1002/(SICI)1097-0029(19980515)41:4<285::AID-JEMT1>3.0.CO;2-Q. Microsc Res Tech. 1998. PMID: 9633946
-
Cross-sectional sample preparation by focused ion beam: a review of ion-sample interaction.Microsc Res Tech. 1996 Nov 1;35(4):320-33. doi: 10.1002/(SICI)1097-0029(19961101)35:4<320::AID-JEMT3>3.0.CO;2-Q. Microsc Res Tech. 1996. PMID: 8987026 Review.
-
Reducing focused ion beam damage to transmission electron microscopy samples.J Electron Microsc (Tokyo). 2004;53(5):451-8. doi: 10.1093/jmicro/dfh080. J Electron Microsc (Tokyo). 2004. PMID: 15582946 Review.
Cited by
-
Nanoscale Cross-Sectional Characterization of Thin Layers in Material Assemblies.Nanomaterials (Basel). 2025 May 30;15(11):840. doi: 10.3390/nano15110840. Nanomaterials (Basel). 2025. PMID: 40497888 Free PMC article. Review.
-
Making the Most of your Electrons: Challenges and Opportunities in Characterizing Hybrid Interfaces with STEM.Mater Today (Kidlington). 2021 Nov;50:100-115. doi: 10.1016/j.mattod.2021.05.006. Epub 2021 Jun 19. Mater Today (Kidlington). 2021. PMID: 35241968 Free PMC article.
LinkOut - more resources
Full Text Sources
Other Literature Sources
Research Materials