Reproducibility and accuracy of echocardiographic measurements of left ventricular parameters using real-time three-dimensional echocardiography
- PMID: 15312875
- DOI: 10.1016/j.jacc.2004.05.050
Reproducibility and accuracy of echocardiographic measurements of left ventricular parameters using real-time three-dimensional echocardiography
Abstract
Objectives: We sought to determine whether assessment of left ventricular (LV) function with real-time (RT) three-dimensional echocardiography (3DE) could reduce the variation of sequential LV measurements and provide greater accuracy than two-dimensional echocardiography (2DE).
Background: Real-time 3DE has become feasible as a standard clinical tool, but its accuracy for LV assessment has not been validated.
Methods: Unselected patients (n = 50; 41 men; age, 64 +/- 8 years) presenting for evaluation of LV function were studied with 2DE and RT-3DE. Test-retest variation was performed by a complete restudy by a separate sonographer within 1 h without alteration of hemodynamics or therapy. Magnetic resonance imaging (MRI) images were obtained during a breath-hold, and measurements were made off-line.
Results: The test-retest variation showed similar measurements for volumes but wider scatter of LV mass measurements with M-mode and 2DE than 3DE. The average MRI end-diastolic volume was 172 +/- 53 ml; LV volumes were underestimated by 2DE (mean difference, -54 +/- 33; p < 0.01) but only slightly by RT-3DE (-4 +/- 29; p = 0.31). Similarly, end-systolic volume by MRI (91 +/- 53 ml) was underestimated by 2DE (mean difference, -28 +/- 28; p < 0.01) and by RT-3DE (mean difference, -3 +/- 18; p = 0.23). Ejection fraction by MRI was similar by 2DE (p = 0.76) and RT-3DE (p = 0.74). Left ventricular mass (183 +/- 50 g) was overestimated by M-mode (mean difference, 68 +/- 86 g; p < 0.01) and 2DE (16 +/- 57; p = 0.04) but not RT-3DE (0 +/- 38 g; p = 0.94). There was good inter- and intra-observer correlation between RT-3DE by two sonographers for volumes, ejection fraction, and mass.
Conclusions: Real-time 3DE is a feasible approach to reduce test-retest variation of LV volume, ejection fraction, and mass measurements in follow-up LV assessment in daily practice.
Similar articles
-
Reconstructed versus real-time 3-dimensional echocardiography: comparison with magnetic resonance imaging.J Am Soc Echocardiogr. 2007 Jul;20(7):862-8. doi: 10.1016/j.echo.2006.12.010. J Am Soc Echocardiogr. 2007. PMID: 17617313
-
Left ventricular volume measurement with echocardiography: a comparison of left ventricular opacification, three-dimensional echocardiography, or both with magnetic resonance imaging.Eur Heart J. 2009 Jan;30(1):98-106. doi: 10.1093/eurheartj/ehn484. Epub 2008 Nov 8. Eur Heart J. 2009. PMID: 18997179
-
Comparison of two- and three-dimensional echocardiography with sequential magnetic resonance imaging for evaluating left ventricular volume and ejection fraction over time in patients with healed myocardial infarction.Am J Cardiol. 2007 Feb 1;99(3):300-6. doi: 10.1016/j.amjcard.2006.08.026. Epub 2006 Dec 1. Am J Cardiol. 2007. PMID: 17261386
-
Imaging the heart failure patient-need for accurate measurements of left ventricular volumes and ejection fraction: the role of three-dimensional and contrast echocardiography.Curr Opin Cardiol. 2016 Sep;31(5):459-68. doi: 10.1097/HCO.0000000000000312. Curr Opin Cardiol. 2016. PMID: 27271926 Review.
-
Reference Values for Real Time Three-Dimensional Echocardiography-Derived Left Ventricular Volumes and Ejection Fraction: Review and Meta-Analysis of Currently Available Studies.Echocardiography. 2015 Dec;32(12):1841-50. doi: 10.1111/echo.12972. Epub 2015 Jun 6. Echocardiography. 2015. PMID: 26053260 Review.
Cited by
-
Biomarkers to Predict Reverse Remodeling and Myocardial Recovery in Heart Failure.Curr Heart Fail Rep. 2016 Oct;13(5):207-218. doi: 10.1007/s11897-016-0303-y. Curr Heart Fail Rep. 2016. PMID: 27726056 Review.
-
Echocardiographic Left Ventricular Mass Estimation: Two-Dimensional Area-Length Method is Superior to M-Mode Linear Method in Swine Models of Cardiac Diseases.J Cardiovasc Transl Res. 2020 Aug;13(4):648-658. doi: 10.1007/s12265-019-09937-7. Epub 2019 Dec 11. J Cardiovasc Transl Res. 2020. PMID: 31828537 Free PMC article.
-
Interventional heart wall motion analysis with cardiac C-arm CT systems.Phys Med Biol. 2014 May 7;59(9):2265-84. doi: 10.1088/0031-9155/59/9/2265. Epub 2014 Apr 15. Phys Med Biol. 2014. PMID: 24731942 Free PMC article.
-
Insights into myocardial mechanics in normal and pathologic states using newer echocardiographic techniques.Curr Heart Fail Rep. 2008 Sep;5(3):143-50. doi: 10.1007/s11897-008-0023-z. Curr Heart Fail Rep. 2008. PMID: 18752764 Review.
-
Pragmatic Method Using Blood Pressure Diaries to Assess Blood Pressure Control.Ann Fam Med. 2016 Jan-Feb;14(1):63-9. doi: 10.1370/afm.1883. Ann Fam Med. 2016. PMID: 26755785 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical