Clinical and radiological results of hydroxyapatite-coated acetabular cups
- PMID: 15316676
- PMCID: PMC3456893
- DOI: 10.1007/s00264-004-0563-0
Clinical and radiological results of hydroxyapatite-coated acetabular cups
Abstract
We prospectively studied a consecutive series of 60 primary total hip arthroplasties in 52 patients. All patients had a hydroxyapatite-coated hemispherical acetabular cup (OCTOFIT). Mean patient age was 52.5 (34-73) years. In 38 hips, a 28-mm and in 22 hips a 32-mm diameter cobalt-chrome femoral head was used. We could follow 53 hips for a mean of 81 (28-125) months. Four cups were revised because of osteolysis, and in another six cups, slight osteolysis without migration was seen. The average annual linear wear rate was 0.05 mm (+/-0.05) for 28-mm heads and 0.16 mm (+/-0.09) for 32-mm heads (p<0.0001). Survivorship analysis predicted a survival rate of 86+/-6.7% at 10 years.
Nous avons étudié de façon prospective une série consécutive de 60 arthroplasties totales primaires de hanche chez 52 malades. Tous les malades avaient une cupule acétabulaire hémisphérique enduite d’hydroxyapatite (OCTOFIT). L’âge moyen des malades était de 52,5 ans (34–73). La tête était en chrome-cobalt, de 28 mm pour 38 hanches et de 32 mm pour 22 hanches. Nous avons suivi 53 hanches pendant une moyenne de 82 mois(28–125). Quatre cupules ont été révisées à cause d’une ostéolyse et pour six autres cupules exixtait une discrète ostéolyse sans migration. L’usure linéaire moyenne annuelle était de 0,05 mm (±0,05) pour les têtes de 28 mm et 0,16 mm (±0,09) pour les têtes de 32 mm (p<0,0001). L’analyse de Kaplan–Meier a prédit un taux de la survie de 86±6,7% à 10 années.
Figures
Similar articles
-
The Harris-Galante Porous acetabular component at intermediate follow-up.Orthopedics. 2001 Aug;24(8):747-51. doi: 10.3928/0147-7447-20010801-16. Orthopedics. 2001. PMID: 11518403
-
A comparison of polyethylene wear in hips with cobalt-chrome or zirconia heads. A prospective, randomised study.J Bone Joint Surg Br. 2001 Jul;83(5):742-50. doi: 10.1302/0301-620x.83b5.10941. J Bone Joint Surg Br. 2001. PMID: 11482320
-
Cementless cups do not increase osteolysis risk in metal-on-metal total hip arthroplasty. Apropos of 106 cases.Orthop Traumatol Surg Res. 2009 Nov;95(7):478-90. doi: 10.1016/j.otsr.2009.07.007. Epub 2009 Oct 3. Orthop Traumatol Surg Res. 2009. PMID: 19801209
-
Early failure of hemispheric hydroxyapatite-coated acetabular cups.Clin Orthop Relat Res. 2006 May;446:233-8. doi: 10.1097/01.blo.0000195923.28480.d4. Clin Orthop Relat Res. 2006. PMID: 16721953
-
High wear rates and extensive osteolysis in 3 types of uncemented total hip arthroplasty: a review of the PCA, the Harris Galante and the Profile/Tri-Lock Plus arthroplasties with a minimum of 12 years median follow-up in 96 hips.Acta Orthop. 2006 Aug;77(4):575-84. doi: 10.1080/17453670610012638. Acta Orthop. 2006. PMID: 16929433 Review.
Cited by
-
Excellent results with cementless total hip arthroplasty and alumina-on-alumina pairing: minimum ten-year follow-up.Int Orthop. 2011 Feb;35(2):195-200. doi: 10.1007/s00264-010-1150-1. Epub 2010 Nov 16. Int Orthop. 2011. PMID: 21079952 Free PMC article. Clinical Trial.