[Transpalpebral tonometry with a digital tonometer in healthy eyes and after penetrating keratoplasty]
- PMID: 15322801
- DOI: 10.1007/s00347-004-1082-5
[Transpalpebral tonometry with a digital tonometer in healthy eyes and after penetrating keratoplasty]
Abstract
Purpose: The Goldmann applanation technique established itself as gold standard for clinical intraocular pressure (IOP) measurement. But the device measures by corneal touch and presupposes corneal conditions within a normal range. So precise determination of IOP is crucial in corneal disorders and after corneal surgery. This pilot study presents tonometry data of normal and postkeratoplasty eyes measured transpalpebrally with a digital tonometer in comparison to Goldmann tonometry.
Method: The TGDc-01 tonometer is based on a ballistic principle of measurement. IOP was measured in 50 healthy eyes (group 1) and 40 eyes after penetrating keratoplasty (group 2) first by the TGDc-01 and after 3 minutes by Goldmann tonometry (Haag-Streit). Subjective refraction, keratometry and ultrasonic pachymetry were performed in all eyes. The IOP measurements in the postkeratoplasty eyes were done when corneal epithelium was fully closed and only a medium suture edema could be observed.
Results: The Pearson's product-moment correlation was r=0,46 for group 1 and 0,65 for group 2. Using the concordance correlation coefficient the value showed a reproducibility index of r(c)=0,44 in group 1 and r(c)=0,60 in group 2. Following the Bland-Altman-analysis 58% of the cases in group 1 had a difference between the two instruments of maximum 3 mmHg. In group 2, 53,5% of the eyes showed an absolute difference between Goldmann and TDGc-01 of 3 mmHg or less.
Conclusion: A transpalpebral measurement technique-avoiding corneal contact-seems to be an alternative choice in patients with corneal pathologies and after corneal surgery. The TGDc-01 tonometer provided mostly similar results compared to the Goldmann device, but with fluctuations. Keeping the deviations between the two methods in mind we recommend, however, the use of the transpalpebral approach only in combination with other techniques (palpation, indentation) to obtain enough reliability of IOP information in problem eyes.
Similar articles
-
Measurement of the intraocular pressure with the "transpalpebral tonometer" TGDc-01 in comparison with applanation tonometry.Graefes Arch Clin Exp Ophthalmol. 2005 Jun;243(6):563-9. doi: 10.1007/s00417-004-1037-1. Epub 2004 Dec 22. Graefes Arch Clin Exp Ophthalmol. 2005. PMID: 15614540 Clinical Trial.
-
[Intraocular pressure difference in Goldmann applanation tonometry versus a transpalpebral tonometer TGDc-01"PRA" in glaucoma patients].Klin Monbl Augenheilkd. 2005 Feb;222(2):123-31. doi: 10.1055/s-2005-857881. Klin Monbl Augenheilkd. 2005. PMID: 15719316 Clinical Trial. German.
-
Measurements of intraocular pressure by Goldmann tonometry, Tonopen XL, and the transpalpebral tonometer, TGDc-01, after penetrating keratoplasty: a comparativye study.Cornea. 2009 Aug;28(7):724-7. doi: 10.1097/ICO.0b013e3181930be8. Cornea. 2009. PMID: 19574919
-
[Measuring intraocular pressure by different methods].Ophthalmologe. 2009 Aug;106(8):676-82. doi: 10.1007/s00347-009-1971-8. Ophthalmologe. 2009. PMID: 19652979 Review. German.
-
Intraocular Pressure Measurement after Penetrating Keratoplasty.Diagnostics (Basel). 2022 Jan 19;12(2):234. doi: 10.3390/diagnostics12020234. Diagnostics (Basel). 2022. PMID: 35204325 Free PMC article. Review.
Cited by
-
Dual Optical Signal-based Intraocular Pressure-sensing Principle Using Pressure-sensitive Mechanoluminescent ZnS:Cu/PDMS Soft Composite.Sci Rep. 2019 Oct 23;9(1):15215. doi: 10.1038/s41598-019-51771-z. Sci Rep. 2019. PMID: 31645632 Free PMC article.
-
Deviations between transpalpebral tonometry using TGDc-01 and Goldmann applanation tonometry depending on the IOP level.Graefes Arch Clin Exp Ophthalmol. 2005 Sep;243(9):853-8. doi: 10.1007/s00417-005-1142-9. Epub 2005 Mar 15. Graefes Arch Clin Exp Ophthalmol. 2005. PMID: 16158308
-
Transpalpebral tonometry: reliability and comparison with Goldmann applanation tonometry and palpation in healthy volunteers.Br J Ophthalmol. 2005 Mar;89(3):280-3. doi: 10.1136/bjo.2004.050211. Br J Ophthalmol. 2005. PMID: 15722303 Free PMC article.
-
[Glaucoma attack following Hughes tarsoconjunctival flap: what should be done?].Ophthalmologe. 2012 Sep;109(9):897-900. doi: 10.1007/s00347-012-2527-x. Ophthalmologe. 2012. PMID: 22740014 German.
-
Central corneal thickness and Diaton transpalpebral tonometry.Graefes Arch Clin Exp Ophthalmol. 2008 Jun;246(6):881-9. doi: 10.1007/s00417-008-0769-8. Epub 2008 Feb 20. Graefes Arch Clin Exp Ophthalmol. 2008. PMID: 18286299
References
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources