Evaluation of accuracy and reproducibility of peak flowmeters at 1,400 m
Abstract
Peak flow meters provide physicians and patients with objective measures about changes in pulmonary obstruction. We evaluated eight models of peak flowmeters and measured their accuracy and reproducibility with methods recently recommended by the National Asthma Education Program (NAEP). Waveforms from the American Thoracic Society's spirometer testing set were used to drive a computer-controlled syringe. Testing was done at Salt Lake City at an altitude 1,400 m. It appears that the original Wright peak flowmeter has been used as the "de facto" standard. We found that the original Wright peak flowmeter overestimated flows in its midrange; and, as a consequence, most of the other peak flowmeters also overestimated peak flows. The overestimation of peak flows may have been understated because of the 1,400-m altitude testing site. To the credit of the instrument manufacturers, we were pleasantly surprised with the quality, accuracy, and reproducibility of presently available peak flowmeters; however, as a result of our testing, we suspect that with little effort, manufacturers of peak flowmeters could improve the accuracy of their devices. Standardized testing methods and equipment should make the task of peak flowmeter design, manufacture, and testing even easier. We trust that manufacturers of peak flowmeters will respond appropriately and improve their instruments.
Similar articles
-
Effect of altitude on hand-held peak flowmeters.Chest. 1996 Feb;109(2):475-9. doi: 10.1378/chest.109.2.475. Chest. 1996. PMID: 8620725
-
Accuracy, reproducibility, and variability of portable peak flowmeters.Chest. 1995 Mar;107(3):648-51. doi: 10.1378/chest.107.3.648. Chest. 1995. PMID: 7874931
-
[An evaluation of three commercially available peak flowmeters].Arerugi. 1994 Apr;43(4):571-8. Arerugi. 1994. PMID: 8031253 Japanese.
-
Measuring peak expiratory flow in general practice: comparison of mini Wright peak flow meter and turbine spirometer.BMJ. 1990 Jun 23;300(6740):1629-31. doi: 10.1136/bmj.300.6740.1629. BMJ. 1990. PMID: 2142611 Free PMC article. Clinical Trial.
-
Selection of peak flowmeters in ambulatory asthma patients: a review of the literature.Chest. 1998 Aug;114(2):587-92. doi: 10.1378/chest.114.2.587. Chest. 1998. PMID: 9726749 Review.
Cited by
-
Care of asthma. A personal approach.Can Fam Physician. 1993 Nov;39:2369-74. Can Fam Physician. 1993. PMID: 8268744 Free PMC article.
-
Potential effects of correction of inaccuracies of the mini-Wright peak expiratory flow meter on the use of an asthma self-management plan.Thorax. 1996 Apr;51(4):403-6. doi: 10.1136/thx.51.4.403. Thorax. 1996. PMID: 8733493 Free PMC article.
-
National Athletic Trainers' Association position statement: management of asthma in athletes.J Athl Train. 2005 Jul-Sep;40(3):224-45. J Athl Train. 2005. PMID: 16284647 Free PMC article.
-
Peak flow meters: a problem of scale.BMJ. 1994 Feb 26;308(6928):548-9. doi: 10.1136/bmj.308.6928.548. BMJ. 1994. PMID: 8148675 Free PMC article. No abstract available.
-
Blow what's the flow?Indian J Pediatr. 1994 Jan-Feb;61(1):110-1. doi: 10.1007/BF02753574. Indian J Pediatr. 1994. PMID: 7927592 No abstract available.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources