Comparative Study
. 2004 Sep;20(3):395-402.
doi: 10.1002/jmri.20141.

Free-breathing, three-dimensional coronary artery magnetic resonance angiography: comparison of sequences

Affiliations
Comparative Study

Free-breathing, three-dimensional coronary artery magnetic resonance angiography: comparison of sequences

Oliver M Weber et al. J Magn Reson Imaging. 2004 Sep.

Abstract

Purpose: To compare six free-breathing, three-dimensional, magnetization-prepared coronary magnetic resonance angiography (MRA) sequences.

Materials and methods: Six bright-blood sequences were evaluated: Cartesian segmented gradient echo (C-SGE), radial SGE (R-SGE), spiral SGE (S-SGE), spiral gradient echo (S-GE), Cartesian steady-state free precession (C-SSFP), and radial SSFP (R-SSFP). The right coronary artery (RCA) was imaged in 10 healthy volunteers using all six sequences in randomized order. Images were evaluated by two observers with respect to signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), visible vessel length, vessel edge sharpness, and vessel diameter.

Results: C-SSFP depicted RCA over the longest distance with high vessel sharpness, good SNR, and excellent background suppression. S-GE provided best SNR and CNR in proximal segments, but more vessel blurring and poorer background suppression, resulting in poor visualization of distal segments. R-SSFP images showed good background suppression and best vessel sharpness, but only moderate SNR. C-SGE provided good SNR and reasonable CNR, but lowest vessel sharpness. S-SGE and R-SGE visualized the RCA over the smallest distance, mostly due to vessel blurring and low SNR, respectively.

Conclusion: Overall, Cartesian SSFP provided the best image quality with excellent vessel sharpness, visualization of long vessel segments, and good SNR and CNR.