Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Clinical Trial
. 2004 Jul-Aug;24(4):359-64.

The self-locating catheter: clinical experience and follow-up

Affiliations
  • PMID: 15335150
Clinical Trial

The self-locating catheter: clinical experience and follow-up

Nicola Di Paolo et al. Perit Dial Int. 2004 Jul-Aug.

Abstract

Background: The self-locating catheter invented by Nicola Di Paolo has been used increasingly in Italy and elsewhere since 1994, with about a thousand patients currently implanted every year. Twelve grams of tungsten inserted into the tip of the conventional Tenckhoff catheter during extrusion does not significantly change its form, but suffices to keep the tip firmly in the Douglas cavity.

Objective: The aim of the present study was to confirm our preliminary results in a large population of peritoneal dialysis patients.

Setting: 16 Italian nephrology departments.

Results: In addition to confirming the validity of the new catheter, the present results show that patients with the new catheter have fewer episodes of peritonitis, tunnel infection, cuff extrusion, catheter malfunction, obstruction, and leakage.

Conclusion: The present multicenter control study confirms preliminary results and demonstrates that complications of peritoneal dialysis, such as cuff extrusion, infection, peritonitis, early leakage, and obstruction, are statistically less frequent in patients with self-locating catheters than in patients with classic Tenckhoff catheters.