Measurements of intrafraction motion and interfraction and intrafraction rotation of prostate by three-dimensional analysis of daily portal imaging with radiopaque markers
- PMID: 15337537
- DOI: 10.1016/j.ijrobp.2004.02.045
Measurements of intrafraction motion and interfraction and intrafraction rotation of prostate by three-dimensional analysis of daily portal imaging with radiopaque markers
Abstract
Purpose: To measure the interfraction and intrafraction motion of the prostate during the course of external beam radiotherapy using a video electronic portal imaging device and three-dimensional analysis.
Methods and materials: Eighteen patients underwent implantation with two or three gold markers in the prostate before five-angle/11-field conformal radiotherapy. Using CT data as the positional reference, multiple daily sets of portal images, and a three-dimensional reconstruction algorithm, intrafraction translations, as well as interfraction and intrafraction rotations, were analyzed along the three principal axes (left-right [LR], superoinferior [SI], and AP). The overall mean values and standard deviations (SDs), along with random and systematic SDs, were computed for these translations and rotations.
Results: For 282 intrafraction translational displacements, the random SD was 0.8 mm (systematic SD, 0.2) in the LR, 1.0 mm (systematic SD, 0.4) in the SI, and 1.4 mm (systematic SD, 0.7) in the AP axes. The analysis of 348 interfraction rotations revealed random SDs of 6.1 degrees (systematic SD, 5.6 degrees ) around the LR axis, 2.8 degrees (systematic SD, 2.4 degrees ) around the SI axis, and 2.0 degrees (systematic SD, 2.2 degrees ) around the AP axis. The intrafraction rotational motion observed during 44 fractions had a random SD of 1.8 degrees (systematic SD, 1.0 degrees ) around the LR, 1.1 degrees (systematic SD, 0.8 degrees ) around the SI, and 0.6 degrees (systematic SD, 0.3 degrees ) around the AP axis.
Conclusion: The interfraction rotations observed were more important than those reported in previous studies. Intrafraction motion was generally smaller in magnitude than interfraction motion. However, the intrafraction rotations and translations of the prostate should be taken into account when designing planning target volume margins because their magnitudes are not negligible.
Similar articles
-
Strategies to reduce the systematic error due to tumor and rectum motion in radiotherapy of prostate cancer.Radiother Oncol. 2005 Feb;74(2):177-85. doi: 10.1016/j.radonc.2004.12.010. Radiother Oncol. 2005. PMID: 15734206 Clinical Trial.
-
A randomized comparison of interfraction and intrafraction prostate motion with and without abdominal compression.Radiother Oncol. 2008 Jul;88(1):88-94. doi: 10.1016/j.radonc.2008.01.019. Epub 2008 Feb 14. Radiother Oncol. 2008. PMID: 18279985 Clinical Trial.
-
Six degrees-of-freedom prostate and lung tumor motion measurements using kilovoltage intrafraction monitoring.Int J Radiat Oncol Biol Phys. 2015 Feb 1;91(2):368-75. doi: 10.1016/j.ijrobp.2014.09.040. Epub 2014 Nov 25. Int J Radiat Oncol Biol Phys. 2015. PMID: 25445555
-
Organ motion and its management.Int J Radiat Oncol Biol Phys. 2001 May 1;50(1):265-78. doi: 10.1016/s0360-3016(01)01453-5. Int J Radiat Oncol Biol Phys. 2001. PMID: 11316572 Review.
-
Radiotherapy of mobile tumors.Semin Radiat Oncol. 2006 Oct;16(4):239-48. doi: 10.1016/j.semradonc.2006.04.007. Semin Radiat Oncol. 2006. PMID: 17010907 Review.
Cited by
-
A treatment planning study comparing IMRT techniques and cyber knife for stereotactic body radiotherapy of low-risk prostate carcinoma.Radiat Oncol. 2019 Aug 9;14(1):143. doi: 10.1186/s13014-019-1353-6. Radiat Oncol. 2019. PMID: 31399115 Free PMC article.
-
The effect of interfraction prostate motion on IMRT plans: a dose-volume histogram analysis using a Gaussian error function model.J Appl Clin Med Phys. 2009 Sep 30;10(4):79-95. doi: 10.1120/jacmp.v10i4.3055. J Appl Clin Med Phys. 2009. PMID: 19918231 Free PMC article.
-
Maintaining prostate contouring consistency following an educational intervention.J Med Radiat Sci. 2016 Sep;63(3):155-60. doi: 10.1002/jmrs.168. Epub 2016 Mar 16. J Med Radiat Sci. 2016. PMID: 27648279 Free PMC article.
-
Dosimetric comparison of image guidance by megavoltage computed tomography versus bone alignment for prostate cancer radiotherapy.Strahlenther Onkol. 2009 Apr;185(4):241-7. doi: 10.1007/s00066-009-1913-z. Epub 2009 Apr 16. Strahlenther Onkol. 2009. PMID: 19370427 Clinical Trial.
-
A novel radiopaque tissue marker for soft tissue localization and in vivo length and area measurements.PLoS One. 2019 Oct 18;14(10):e0224244. doi: 10.1371/journal.pone.0224244. eCollection 2019. PLoS One. 2019. PMID: 31626672 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical