Prediction of fixation failure after sliding hip screw fixation
Abstract
Cut-out of the lag screw is the commonest cause of fixation failure after sliding hip screw fixation of extracapsular hip fracture. A number of technical aspects of surgery have been used to asses the risk of cut-out. This study was to determine which of these indicators was the most reliable predictor of cut-out. The anterior-posterior and lateral post-operative radiographs of 23 cases of cut-out were compared with those of 77 cases of uneventful fracture healing. The tip-apex distance with correction for magnification was found to show the most significant difference between patients with cut-out against those without (P = 0.001), followed by the lag screw position on the lateral radiographs (P = 0.0095 and 0.014), reduction of the fracture on the anterior-posterior radiograph (P = 0.011 and 0.016) and the uncorrected tip-apex distance (P = 0.019). We recommend that for audit and research purposes the corrected tip-apex distance, fracture reduction and implant positioning methods should be used. For routine clinical practice, the uncorrected tip to apex distance, which is sum of the distance from the tip of the lag screw to the apex of the femoral head on anterior-posterior and lateral radiograph, and fracture reduction angle on the anterior-posterior radiograph are recommended.
Comment in
-
Prediction of fixation failure after sliding hip screw fixation [Injury 2004;35:994-8].Injury. 2005 Aug;36(8):984-5; author reply 985. doi: 10.1016/j.injury.2004.12.042. Epub 2005 Mar 31. Injury. 2005. PMID: 16023910 No abstract available.
-
Prediction of fixation failure after sliding hip screw fixation.Injury. 2005 Dec;36(12):1491. doi: 10.1016/j.injury.2005.07.043. Epub 2005 Oct 20. Injury. 2005. PMID: 16242691 No abstract available.
Similar articles
-
Risk factors for cut-out of double lag screw fixation in proximal femoral fractures.Injury. 2017 Feb;48(2):414-418. doi: 10.1016/j.injury.2016.11.018. Epub 2016 Nov 19. Injury. 2017. PMID: 27889112
-
A mathematical simulation of the tip-apex distance and the calcar-referenced tip-apex distance for intertrochanteric fractures reduced with lag screws.Injury. 2016 Jun;47(6):1302-8. doi: 10.1016/j.injury.2016.03.029. Epub 2016 Apr 1. Injury. 2016. PMID: 27087281
-
Review of methods to quantify lag screw placement in hip fracture fixation.Acta Orthop Belg. 2005 Jun;71(3):260-3. Acta Orthop Belg. 2005. PMID: 16035697 Review.
-
130- versus 135-degree sliding hip screws and failure in pertrochanteric hip fractures.ANZ J Surg. 2014 Dec;84(12):949-54. doi: 10.1111/ans.12713. Epub 2014 Jun 5. ANZ J Surg. 2014. PMID: 24898151
-
Comparison of tip apex distance and cut-out complications between helical blades and lag screws in intertrochanteric fractures among the elderly: a meta-analysis.J Orthop Sci. 2015 Nov;20(6):1062-9. doi: 10.1007/s00776-015-0770-0. J Orthop Sci. 2015. PMID: 26481124 Review.
Cited by
-
Correlation between cephalic screw positioning of Standard Gamma 3 Nail for intertrochanteric fractures and cut-out incidence.SICOT J. 2024;10:9. doi: 10.1051/sicotj/2024006. Epub 2024 Feb 26. SICOT J. 2024. PMID: 38415767 Free PMC article.
-
Critical factors in cut-out complication after Gamma Nail treatment of proximal femoral fractures.BMC Musculoskelet Disord. 2013 Jan 2;14:1. doi: 10.1186/1471-2474-14-1. BMC Musculoskelet Disord. 2013. PMID: 23281775 Free PMC article.
-
Unstable intertrochanteric femur fractures: is there a consensus on definition and treatment in Germany?Clin Orthop Relat Res. 2013 Sep;471(9):2831-40. doi: 10.1007/s11999-013-2834-9. Clin Orthop Relat Res. 2013. PMID: 23389806 Free PMC article.
-
Circumferential wires as a supplement to intramedullary nailing in unstable trochanteric hip fractures: 4 reoperations in 60 patients followed for 1 year.Acta Orthop. 2012 Jun;83(3):240-3. doi: 10.3109/17453674.2012.665329. Epub 2012 Feb 13. Acta Orthop. 2012. PMID: 22329672 Free PMC article.
-
Anchorage strategies in geriatric hip fracture management.Innov Surg Sci. 2016 Dec 22;1(2):73-78. doi: 10.1515/iss-2016-0034. eCollection 2016 Dec. Innov Surg Sci. 2016. PMID: 31579722 Free PMC article. Review.