A dual closed-loop control system for mechanical ventilation
- PMID: 15362273
- DOI: 10.1023/b:jocm.0000032744.99885.38
A dual closed-loop control system for mechanical ventilation
Abstract
Objective: Closed-loop mechanical ventilation has the potential to provide more effective ventilatory support to patients with less complexity than conventional ventilation. The purpose of this study was to investigate the effectiveness of an automatic technique for mechanical ventilation.
Methods: Two closed-loop control systems for mechanical ventilation are combined in this study. In one of the control systems several physiological data are used to automatically adjust the frequency and tidal volume of breaths of a patient. This method, which is patented under US Patent number 4986268, uses the criterion of minimal respiratory work rate to provide the patient with a natural pattern of breathing. The inputs to the system include data representing CO2 and O2 levels of the patient as well as respiratory compliance and airway resistance. The I:E ratio is adjusted on the basis of the respiratory time constant to allow for effective emptying of the lungs in expiration and to avoid intrinsic positive end expiratory pressure (PEEP). This system is combined with another closed-loop control system for automatic adjustment of the inspired fraction of oxygen of the patient. This controller uses the feedback of arterial oxygen saturation of the patient and combines a rapid stepwise control procedure with a proportional-integral-derivative (PID) control algorithm to automatically adjust the oxygen concentration in the patient's inspired gas. The dual closed-loop control system has been examined by using mechanical lung studies, computer simulations and animal experiments.
Results: In the mechanical lung studies, the ventilation controller adjusted the breathing frequency and tidal volume in a clinically appropriate manner in response to changes in respiratory mechanics. The results of computer simulations and animal studies under induced disturbances showed that blood gases were returned to the normal physiologic range in less than 25 s by the control system. In the animal experiments understeady-state conditions, the maximum standard deviations of arterial oxygen saturation and the end-tidal partial pressure of CO2 were +/- 1.76% and +/- 1.78 mmHg, respectively.
Conclusion: The controller maintained the arterial blood gases within normal limits under steady-state conditions and the transient response of the system was robust under various disturbances. The results of the study have showed that the proposed dual closed-loop technique has effectively controlled mechanical ventilation under different test conditions.
Similar articles
-
Closed-loop control if the inspired fraction of oxygen in mechanical ventilation.J Clin Monit Comput. 2002 Aug;17(6):367-76. doi: 10.1023/a:1024261021473. J Clin Monit Comput. 2002. PMID: 12885181
-
[Adaptive lung ventilation (AVL). Evaluation of new closed loop regulated respiration algorithm for operation in the hyperextended lateral position].Anaesthesist. 1996 Oct;45(10):950-6. doi: 10.1007/s001010050329. Anaesthesist. 1996. PMID: 8992909 Clinical Trial. German.
-
Automatic selection of tidal volume, respiratory frequency and minute ventilation in intubated ICU patients as start up procedure for closed-loop controlled ventilation.Int J Clin Monit Comput. 1994 Feb;11(1):19-30. doi: 10.1007/BF01132840. Int J Clin Monit Comput. 1994. PMID: 8195655
-
Enhancement of mechanical ventilation of neonates by computer technology.Semin Perinatol. 2000 Dec;24(6):429-44. doi: 10.1053/sper.2000.20084. Semin Perinatol. 2000. PMID: 11153904 Review.
-
Closed-loop mechanical ventilation.Respir Care. 2002 Apr;47(4):427-51; discussion 451-3. Respir Care. 2002. PMID: 11929616 Review.
Cited by
-
Newer nonconventional modes of mechanical ventilation.J Emerg Trauma Shock. 2014 Jul;7(3):222-7. doi: 10.4103/0974-2700.136869. J Emerg Trauma Shock. 2014. PMID: 25114434 Free PMC article. Review.
-
Flex: a new computerized system for mechanical ventilation.J Clin Monit Comput. 2008 Apr;22(2):121-30. doi: 10.1007/s10877-008-9113-4. Epub 2008 Mar 7. J Clin Monit Comput. 2008. PMID: 18324476
-
Robustness of two different methods of monitoring respiratory system compliance during mechanical ventilation.Med Biol Eng Comput. 2017 Oct;55(10):1819-1828. doi: 10.1007/s11517-017-1631-0. Epub 2017 Feb 27. Med Biol Eng Comput. 2017. PMID: 28243966 Free PMC article.
-
Evaluation of a computerized system for mechanical ventilation of infants.J Clin Monit Comput. 2009 Apr;23(2):93-104. doi: 10.1007/s10877-009-9170-3. Epub 2009 Mar 5. J Clin Monit Comput. 2009. PMID: 19263230
-
Adaptive support ventilation is a patented technology.Intensive Care Med. 2008 Dec;34(12):2306; author reply 2307-8. doi: 10.1007/s00134-008-1290-6. Epub 2008 Sep 25. Intensive Care Med. 2008. PMID: 18815770 No abstract available.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources