Clinical Trial
. 2004 Sep;30(9):1776-82.
doi: 10.1007/s00134-004-2380-8. Epub 2004 Jul 28.

Bedside estimation of absolute renal blood flow and glomerular filtration rate in the intensive care unit. A validation of two independent methods

Affiliations
Clinical Trial

Bedside estimation of absolute renal blood flow and glomerular filtration rate in the intensive care unit. A validation of two independent methods

Kristina Swärd et al. Intensive Care Med. 2004 Sep.

Abstract

Objective: To evaluate various treatment strategies in critically ill patients with ischaemic acute renal failure, there is a need for reliable bedside measurements of total renal blood flow (RBF), glomerular filtration rate (GFR) and renal oxygen consumption without the need for urine collection.

Design: The continuous renal vein thermodilution method and the infusion clearance techniques were validated against the gold standard technique, the urinary clearance of paraaminohippurate (PAH) and chromium ethylenediaminetetraacetic acid, respectively.

Setting: University hospital cardiothoracic ICU.

Patients: Seventeen uncomplicated mechanically ventilated post-cardiac surgical patients.

Interventions: None.

Measurements and results: Renal blood flow, GFR and the renal filtration fraction (FF) were measured for two consecutive 30-min periods by urinary clearance and compared with simultaneous measurements made by the thermodilution and infusion clearance techniques. Urinary clearance for PAH was corrected for by renal extraction of PAH. The within-group error, repeatability coefficient and the coefficient of variation were highest for the thermodilution technique and lowest for the infusion clearance technique with regard to RBF, GFR and FF. The infusion clearance technique had a higher agreement with the urinary clearance method than the thermodilution method. For estimations of RBF and GFR, the between-group errors were 33% and 43% comparing infusion clearance with urinary clearance and 65% and 67% comparing thermodilution with urinary clearance.

Conclusions: The infusion clearance method had the highest reproducibility and the highest agreement with the urinary clearance reference method. The renal vein thermodilution technique is less reliable in the ICU setting due to poor repeatability and poor agreement with the reference method.

PubMed Disclaimer

Similar articles

Cited by

References

    1. Acta Physiol Scand. 1988 Sep;134(1):23-34 - PubMed
    1. Clin Physiol. 1988 Aug;8(4):407-16 - PubMed
    1. Crit Care Med. 1992 May;20(5):659-64 - PubMed
    1. Acta Anaesthesiol Scand. 2001 May;45(5):536-42 - PubMed
    1. Chest. 1990 Jul;98(1):170-9 - PubMed

Publication types

LinkOut - more resources