Reflection and transmission pulse oximetry during compromised peripheral perfusion
- PMID: 1538246
- DOI: 10.1007/BF01618081
Reflection and transmission pulse oximetry during compromised peripheral perfusion
Abstract
The performance of a reflection pulse oximeter and a transmission pulse oximeter was assessed during open-heart surgery when cardiac output, peripheral temperature, pulse pressure, and systolic pressure were low and vascular resistance was high. Before and after extracorporeal circulation (ECC) there was no difference in ability of the sensors to obtain readings and no difference in the accuracy of those readings. During partial ECC, especially after coronary artery bypass grafting, the reflection sensor gave readings earlier and at a lower pulse pressure. In addition, the transmission sensor failed to give any readings for 2 patients on partial ECC, for whom the reflection sensor did give readings. The accuracy of heart rate (HR) data was comparable for both sensors before ECC; however, during partial ECC, the reflection sensor tended to give values closer to the electrocardiographic HR. The accuracy of saturation data given by the reflection oximeter was comparable to that of the transmission oximeter. It is concluded that the accuracy of the saturation and HR data provided by the two methods of pulse oximetry are comparable, but that the reflection sensor is more likely to obtain readings under conditions of poorer peripheral circulation.
Similar articles
-
Continuous intra-arterial oximetry, pulse oximetry, and co-oximetry during cardiac surgery.J Cardiothorac Vasc Anesth. 1992 Dec;6(6):668-73. doi: 10.1016/1053-0770(92)90048-c. J Cardiothorac Vasc Anesth. 1992. PMID: 1472661
-
Comparison of Pulse Oximetry and Handheld Doppler for Assessing Digital Perfusion.Bull Hosp Jt Dis (2013). 2019 Mar;77(2):128-131. Bull Hosp Jt Dis (2013). 2019. PMID: 31128582
-
Effect of peripheral perfusion on accuracy of pulse oximetry in children.J Clin Anesth. 1999 Jun;11(4):317-22. doi: 10.1016/s0952-8180(99)00054-9. J Clin Anesth. 1999. PMID: 10470634
-
Pulse oximetry in the oesophagus.Physiol Meas. 2006 Jan;27(1):R1-35. doi: 10.1088/0967-3334/27/1/R01. Epub 2005 Nov 29. Physiol Meas. 2006. PMID: 16365505 Review.
-
Accuracy of fetal pulse oximetry and pitfalls in measurements.Eur J Obstet Gynecol Reprod Biol. 1997 Mar;72 Suppl:S21-7. doi: 10.1016/s0301-2115(97)02714-0. Eur J Obstet Gynecol Reprod Biol. 1997. PMID: 9134409 Review.
Cited by
-
Pulse oximetry and high-dose vasopressors: a comparison between forehead reflectance and finger transmission sensors.Intensive Care Med. 2012 Oct;38(10):1718-22. doi: 10.1007/s00134-012-2659-0. Epub 2012 Aug 7. Intensive Care Med. 2012. PMID: 22868275
-
The influence of larger subcutaneous blood vessels on pulse oximetry.J Clin Monit Comput. 2004 Jun;18(3):179-88. doi: 10.1023/b:jocm.0000042924.16052.5e. J Clin Monit Comput. 2004. PMID: 15562984
-
Reflection-Boosted Wearable Ring-Type Pulse Oximeters for SpO2 Measurement with High Sensitivity and Low Power Consumption.Biosensors (Basel). 2023 Jul 5;13(7):711. doi: 10.3390/bios13070711. Biosensors (Basel). 2023. PMID: 37504110 Free PMC article.
-
Reflectance pulse oximetry at the forehead improves by pressure on the probe.J Clin Monit. 1995 Jul;11(4):237-44. doi: 10.1007/BF01617518. J Clin Monit. 1995. PMID: 7561997
-
Effect of functional neuromuscular stimulation on postural related orthostatic stress in individuals with acute spinal cord injury.J Rehabil Res Dev. 2000 Sep-Oct;37(5):535-42. J Rehabil Res Dev. 2000. PMID: 11322152 Free PMC article. Clinical Trial.