A comparison of bispectral index and rapidly extracted auditory evoked potentials index responses to noxious stimulation during sevoflurane anesthesia
- PMID: 15385365
- DOI: 10.1213/01.ANE.0000130618.99860.48
A comparison of bispectral index and rapidly extracted auditory evoked potentials index responses to noxious stimulation during sevoflurane anesthesia
Abstract
In 21 patients given sevoflurane anesthesia, we simultaneously compared the abilities of Bispectral Index (BIS) and rapidly extracted auditory evoked potentials index (AAI) to display the effect of an increasing cerebral concentration of sevoflurane, with and without noxious stimulation. In addition to BIS/AAI, hemodynamic variables were monitored. After titrating sevoflurane to BIS = 50-55 during 15 min, the end-tidal concentration of sevoflurane (1.46% +/- 0.20%) was doubled followed by a noxious stimulus, laryngoscopy, applied at random time points within the following 15 min. After the end-tidal concentration of sevoflurane was doubled, a substantial reduction in BIS was observed, whereas only a slight reduction in AAI was seen (P < 0.0001). BIS/AAI responses to laryngoscopy were not attenuated with increasing wash-in of sevoflurane. After noxious stimulation, AAI exceeded the highest recommended value, 25, in 3 cases, whereas BIS did not exceed the recommended threshold, 60, in any of the patients. Response times for BIS and AAI were 44.5 +/- 26 and 47 +/- 31 s, respectively. These results suggest that, at a hypnotic level associated with surgical sevoflurane anesthesia, BIS better displays drug-related alterations in the level of hypnosis than AAI or hemodynamic variables but there is no difference between BIS and AAI in the time to response to a noxious stimulus.
References
-
- Sandin RH, Enlund G, Samuelsson P, Lennmarken C. Awareness during anaesthesia: a prospective case study. Lancet 2000;355:707–11.
-
- Rampil IJ. A primer for EEG signal processing in anesthesia potency. Anesthesiology 1998;89:980–1002.
-
- Urhonen E, Jensen EW, Lund J. Changes in rapidly extracted auditory evoked potentials during tracheal intubation. Acta Anaesthesiol Scand 2000;44:743–8.
-
- Schneider G, Wagner K, Reeker W, et al. Bispectral index (BIS) may not predict awareness reaction to intubation in surgical patients. J Neurosurg Anesthesiol 2002;14:7–11.
-