Determination of cardiac output by an angle and diameter independent dual beam Doppler technique in critically ill infants
- PMID: 1540440
- PMCID: PMC1024751
- DOI: 10.1136/hrt.67.2.180
Determination of cardiac output by an angle and diameter independent dual beam Doppler technique in critically ill infants
Abstract
Objective: To compare cardiac output measurements in critically ill infants by the dual beam Doppler and thermodilution techniques.
Design: Prospective direct comparison of the two techniques. For statistical evaluation one randomly assigned paired measurement of every patient was used.
Setting: Paediatric intensive care unit in a university hospital.
Patients: 18 infants after open heart surgery aged 4-25 months (weight 4-10 kg).
Interventions: Cardiac output measurements by dual beam Doppler and thermodilution techniques were performed within 10 minutes of each other and without knowledge of the results of the other methods. Multiple measurements were performed on some patients with a pharmacological or electrophysiological intervention or with a minimum of six hours between each pair of measurements.
Measurements and main results: Three patients were excluded because of an inadequate Doppler signal or a significant residual shunt. Cardiac output measurements ranged from 0.4 to 2.2 l/min for the thermodilution technique and from 0.5 to 2.1 l/min for the dual beam Doppler technique. Agreement between both methods was acceptable. The mean difference between the two methods was 0.026 l/min with two standard deviations ranging from -0.20 to 0.26 l/min.
Conclusion: The dual beam Doppler technique was shown to have promise for the non-invasive determination of cardiac output in critically ill infants.
Similar articles
-
Determination of cardiac output in critically ill patients by dual beam Doppler echocardiography.J Am Coll Cardiol. 1989 Feb;13(2):340-7. doi: 10.1016/0735-1097(89)90509-3. J Am Coll Cardiol. 1989. PMID: 2913112
-
A comparison of thermodilution and pulsed Doppler cardiac output measurement in critically ill children.J Pediatr. 1989 Oct;115(4):554-60. doi: 10.1016/s0022-3476(89)80280-x. J Pediatr. 1989. PMID: 2795345
-
Transoesophageal Doppler echocardiographic measurement of cardiac output by the mitral annulus method.Br Heart J. 1992 Nov;68(5):510-5. doi: 10.1136/hrt.68.11.510. Br Heart J. 1992. PMID: 1467040 Free PMC article.
-
Continuous measurement of cardiac output by the Fick principle in infants and children: comparison with the thermodilution method.Intensive Care Med. 1996 May;22(5):467-71. doi: 10.1007/BF01712169. Intensive Care Med. 1996. PMID: 8796404 Review.
-
The validity of trans-esophageal Doppler ultrasonography as a measure of cardiac output in critically ill adults.Intensive Care Med. 2004 Nov;30(11):2060-6. doi: 10.1007/s00134-004-2430-2. Epub 2004 Sep 11. Intensive Care Med. 2004. PMID: 15368037 Review.
Cited by
-
Accuracy and precision of minimally-invasive cardiac output monitoring in children: a systematic review and meta-analysis.J Clin Monit Comput. 2016 Oct;30(5):603-20. doi: 10.1007/s10877-015-9757-9. Epub 2015 Aug 29. J Clin Monit Comput. 2016. PMID: 26315477
-
Accuracy and repeatability of pediatric cardiac output measurement using Doppler: 20-year review of the literature.Intensive Care Med. 2003 Nov;29(11):1889-94. doi: 10.1007/s00134-003-1967-9. Epub 2003 Sep 4. Intensive Care Med. 2003. PMID: 12955181 Review.
References
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical