Hypoxaemia is reduced by pulse oximetry monitoring in the operating theatre and in the recovery room
- PMID: 1540455
- DOI: 10.1093/bja/68.2.146
Hypoxaemia is reduced by pulse oximetry monitoring in the operating theatre and in the recovery room
Abstract
To determine the impact of pulse oximeter monitoring on the incidence, severity and duration of hypoxaemia in the operating theatre (OT) and in the recovery room (RR), we investigated 200 patients in a randomized study. The extent of hypoxaemia in the OT was compared with that in the RR. Adult inpatients were allocated randomly to two groups: group I, pulse oximeter data and alarms "available"; group II, these data "unavailable" to the anaesthesia team and RR staff. Hypoxaemia was graded into four values of oxyhaemoglobin saturation (SpO2). The incidence of hypoxaemia was reduced significantly in group I in both OT and RR. In the OT, five patients in group II suffered SpO2 less than 76% compared with none in group I (P less than 0.02). In group II in the RR, seven patients suffered SpO2 less than 81%; three of these had SpO2 less than 76%. No patients in group I exhibited such small values of saturation. The smallest recorded SpO2 in the OT and the RR was significantly greater in group I. The cumulative duration of hypoxaemia was significantly less in group I in the RR, but not in the OT. The incidence and severity of hypoxaemia in the OT and in the RR were comparable, whereas the cumulative duration of hypoxaemia was significantly greater in the RR than in the OT. The occurrence of hypoxaemia in an individual patient in the OT significantly increased this patient's risk of suffering hypoxaemia in the RR. We conclude that the extent of hypoxaemia, especially in the RR, may be reduced significantly by pulse oximeter monitoring, but even with the information provided, some patients still develop hypoxaemia.
Similar articles
-
Influence of motivation of care providers on the incidence of postoperative hypoxaemia in the recovery room.Br J Anaesth. 1996 Oct;77(4):453-7. doi: 10.1093/bja/77.4.453. Br J Anaesth. 1996. PMID: 8942327
-
Anesthesia related hypoxemia. The effect of pulse oximetry monitoring on perioperative events and postoperative complications.Dan Med Bull. 1994 Nov;41(5):489-500. Dan Med Bull. 1994. PMID: 7859517 Clinical Trial.
-
Hypoxaemia during anaesthesia--an observer study.Br J Anaesth. 1991 Apr;66(4):437-44. doi: 10.1093/bja/66.4.437. Br J Anaesth. 1991. PMID: 2025470
-
Hypoxaemia in adults in the post-anaesthesia care unit.Can J Anaesth. 1991 Sep;38(6):740-6. doi: 10.1007/BF03008452. Can J Anaesth. 1991. PMID: 1914057 Review.
-
[Pulse oximetry in perioperative monitoring].Ugeskr Laeger. 2003 Oct 13;165(42):4025-7. Ugeskr Laeger. 2003. PMID: 14610837 Review. Danish.
Cited by
-
A Comparison of Two Transport Monitor Systems With Regard to Efficiency and Staff Satisfaction in the Perioperative Setting.Cureus. 2024 May 17;16(5):e60481. doi: 10.7759/cureus.60481. eCollection 2024 May. Cureus. 2024. PMID: 38883109 Free PMC article.
-
Avatar-Based Patient Monitoring With Peripheral Vision: A Multicenter Comparative Eye-Tracking Study.J Med Internet Res. 2019 Jul 17;21(7):e13041. doi: 10.2196/13041. J Med Internet Res. 2019. PMID: 31317870 Free PMC article.
-
Is pulse oximetry an essential tool or just another distraction? The role of the pulse oximeter in modern anesthesia care.J Clin Monit Comput. 2013 Jun;27(3):235-42. doi: 10.1007/s10877-013-9428-7. Epub 2013 Jan 12. J Clin Monit Comput. 2013. PMID: 23314807 Review.
-
Post-laparotomy Hypoxia: A Case Series.Cureus. 2022 Aug 17;14(8):e28096. doi: 10.7759/cureus.28096. eCollection 2022 Aug. Cureus. 2022. PMID: 36127966 Free PMC article.
-
The incidence of hypoxemia during surgery: evidence from two institutions.Can J Anaesth. 2010 Oct;57(10):888-97. doi: 10.1007/s12630-010-9366-5. Epub 2010 Jul 31. Can J Anaesth. 2010. PMID: 20680710 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical