[A comparison of two methods for continuous cardiac output measurement: PulseCO VS CCO]
- PMID: 15446687
[A comparison of two methods for continuous cardiac output measurement: PulseCO VS CCO]
Abstract
Background: PulseCO (PulseCO) (PulseCO Hemodynamic Monitor, LiDCO Co., London, England) is a low invasive apparatus to measure cardiac output continuously from arterial pulse waveform. CCO (774 HF 75, Edwards Lifescience Co., California, USA) is a continuous cardiac monitor commonly used clinically. The purpose of this study is to compare the accuracy of these two methods for cardiac output measurement with the thermodilution technique (TDCO) as control.
Methods: To compare the accuracy of PulseCO with that of CCO, six patients with pulmonary-artery catheter inserted were recruited. PulseCO and CCO were measured continuously, and these CO values were compared with TDCO measurements every hour.
Results: Correlation with TDCO was examined in PulseCO (r=0.82) and CCO (r=0.80).
Conclusions: PulseCO was low invasive, and showed a significantly better correlation with TDCO, compared with CCO.
Similar articles
-
Reliability of continuous cardiac output measurement during intra-abdominal hypertension relies on repeated calibrations: an experimental animal study.Crit Care. 2008;12(5):R132. doi: 10.1186/cc7102. Epub 2008 Oct 29. Crit Care. 2008. PMID: 18957114 Free PMC article.
-
[Comparison of cardiac output measured by pulseCO and that by vigilance during inferior vena cava clamping].Masui. 2005 Oct;54(10):1138-42. Masui. 2005. PMID: 16231769 Japanese.
-
[Comparison between continuous and intermittent thermodilution measurement of cardiac output during coronary artery bypass operation].Anasthesiol Intensivmed Notfallmed Schmerzther. 1997 Apr;32(4):226-33. doi: 10.1055/s-2007-995042. Anasthesiol Intensivmed Notfallmed Schmerzther. 1997. PMID: 9289022 German.
-
Cardiac output monitoring.Curr Opin Anaesthesiol. 2009 Feb;22(1):71-7. doi: 10.1097/ACO.0b013e32831f44d0. Curr Opin Anaesthesiol. 2009. PMID: 19295295 Review.
-
Lithium dilution measurement of cardiac output and arterial pulse waveform analysis: an indicator dilution calibrated beat-by-beat system for continuous estimation of cardiac output.Curr Opin Crit Care. 2002 Jun;8(3):257-61. doi: 10.1097/00075198-200206000-00010. Curr Opin Crit Care. 2002. PMID: 12386506 Review.
Cited by
-
Cardiac index value measurement by invasive, semi-invasive and non invasive techniques: a prospective study in postoperative off pump coronary artery bypass surgery patients.J Clin Monit Comput. 2009 Jun;23(3):175-80. doi: 10.1007/s10877-009-9179-7. Epub 2009 May 2. J Clin Monit Comput. 2009. PMID: 19412657 Clinical Trial.
-
Real-time stroke volume measurements for the optimization of cardiac resynchronization therapy parameters.Europace. 2010 Sep;12(9):1270-4. doi: 10.1093/europace/euq175. Epub 2010 Jun 4. Europace. 2010. PMID: 20525728 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Research Materials
Miscellaneous