Comparison of the sensitivities and viral load values of the AMPLICOR HIV-1 MONITOR version 1.0 and 1.5 tests
- PMID: 15528752
- PMCID: PMC525204
- DOI: 10.1128/JCM.42.11.5392-5393.2004
Comparison of the sensitivities and viral load values of the AMPLICOR HIV-1 MONITOR version 1.0 and 1.5 tests
Abstract
This study compared the sensitivity and viral load values of the AMPLICOR HIV-1 MONITOR microwell version 1.0, microwell version 1.5, and COBAS version 1.5 tests. Based on the percentage of positive replicates, the microwell version 1.5 and COBAS version 1.5 tests are more sensitive than the microwell version 1.0 test. Viral load values obtained with the COBAS version 1.5 test are lower than those obtained with either the microwell version 1.0 or microwell version 1.5 test.
Figures
References
-
- Finney, D. J. 1971. Probit analysis, 3rd ed. Cambridge University Press, London, England.
-
- Jagodzinski, L. L., D. L. Wiggins, J. L. McManis, S. Emery, J. Overbaugh, M. Robb, S. Bodrug, and N. L. Michael. 2000. Use of calibrated viral load standards for group M subtypes of human immunodeficiency virus type 1 to assess the performance of viral RNA quantitation tests. J. Clin. Microbiol. 38:1247-1249. - PMC - PubMed
-
- Kirkwood, T. B. L. 1979. Geometric means and measures of dispersion. Biometrics 35:908-909.
-
- Parekh, B., S. Phillips, T. C. Granade, J. Baggs, D. J. Hu, and R. Respess. 1999. Impact of HIV type 1 subtype variation on viral RNA quantitation. AIDS Res. Hum. Retrovir. 15:133-142. - PubMed