Incidence of stent under-deployment as a cause of in-stent restenosis in long stents
- PMID: 15529909
- DOI: 10.1023/b:caim.0000041947.11384.93
Incidence of stent under-deployment as a cause of in-stent restenosis in long stents
Abstract
Although stent under-deployment (SU) has been associated with increased risk of in-stent restenosis, little data have been reported on the incidence of SU in patients presenting with clinical in-stent restenosis. In 59 patients referred for vascular brachytherapy and showing angiographic in-stent restenosis, we sought (1) to determine the incidence of SU using standard intravascular ultrasound (IVUS) criteria (2) to evaluate the effects of repeat angioplasty on further stent expansion. Stented length was 32+/-17 mm and diameter stenosis was 75+/-14%. Before re-intervention, the incidence of reduced absolute values of minimal stent cross-sectional area (MSCSA) varied from 69% (< or =8 mm2) to 15% (< or =5 mm2). After re-intervention, the incidence decreased to 24% (< or =8 mm2) (p = 0.0001) and 0% (< or =5 mm2) (p = 0.005). Before re-intervention, SU as assessed by relative criteria varied from 21% (80% mean reference lumen area or 90% minimum distal reference lumen area) to 28% (100% minimum reference lumen area). After re-intervention, the incidence of SU varied from 7% (90% minimum distal reference lumen area) (p = 0.0001 vs. pre) to 24% (55% mean reference EEM area) (p = ns). No change in strut apposition (97% pre vs. 100% post) nor in symmetry index (100% pre vs. post) was noted. From all criteria, the 90 and 100% minimum reference lumen area criteria were the most altered by repeat balloon dilatation, 21% pre vs. 7% post and 28% pre vs. 11% post, respectively. In conclusion, among patients presenting with severe angiographic in-stent restenosis, a significant number showed signs of SU whose incidence varied according to applied criteria. Significant stent re-expansion can be obtained following IVUS-guided repeat angioplasty irrespective of initial SU criteria.
Comment in
-
Incidence of stent under-deployment as a cause of in-stent restenosis in long stents: take your lesson!Int J Cardiovasc Imaging. 2004 Dec;20(6):529-30. doi: 10.1007/s10554-004-1631-7. Int J Cardiovasc Imaging. 2004. PMID: 15856637 No abstract available.
Similar articles
-
Standardized angiographically guided over-dilatation of stents using high pressure technique optimize results without increasing risks.J Invasive Cardiol. 2002 May;14(5):221-6. J Invasive Cardiol. 2002. PMID: 11983940 Clinical Trial.
-
Repeat stenting for the prevention of the early lumen loss phenomenon in patients with in-stent restenosis. Angiographic and intravascular ultrasound findings of a randomized study.Am Heart J. 2005 Feb;149(2):e1-8. doi: 10.1016/j.ahj.2004.06.022. Am Heart J. 2005. PMID: 15846250 Clinical Trial.
-
Angiography-guided routine coronary stent implantation results in suboptimal dilatation.Angiology. 2002 Jan-Feb;53(1):69-75. doi: 10.1177/000331970205300109. Angiology. 2002. PMID: 11863311
-
Digital enhancement of stent images in primary and secondary percutaneous coronary revascularisation.EuroIntervention. 2009 May;5 Suppl D:D101-6. EuroIntervention. 2009. PMID: 19736057 Review.
-
Endovascular stent-induced alterations in host artery mechanical environments and their roles in stent restenosis and late thrombosis.Regen Biomater. 2018 Jun;5(3):177-187. doi: 10.1093/rb/rby006. Epub 2018 May 2. Regen Biomater. 2018. PMID: 29942650 Free PMC article. Review.
Cited by
-
Conformability of balloon-expandable stents to the carotid siphon: an in vitro study.AJNR Am J Neuroradiol. 2006 Feb;27(2):324-6. AJNR Am J Neuroradiol. 2006. PMID: 16484402 Free PMC article.
-
Physiologic Assessment after Coronary Stent Implantation.Korean Circ J. 2021 Mar;51(3):189-201. doi: 10.4070/kcj.2020.0548. Korean Circ J. 2021. PMID: 33655719 Free PMC article. Review.
-
Incidence of stent under-deployment as a cause of in-stent restenosis in long stents: take your lesson!Int J Cardiovasc Imaging. 2004 Dec;20(6):529-30. doi: 10.1007/s10554-004-1631-7. Int J Cardiovasc Imaging. 2004. PMID: 15856637 No abstract available.
References
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Miscellaneous