Comparison of inverse planning simulated annealing and geometrical optimization for prostate high-dose-rate brachytherapy
- PMID: 15533807
- DOI: 10.1016/j.brachy.2004.05.007
Comparison of inverse planning simulated annealing and geometrical optimization for prostate high-dose-rate brachytherapy
Abstract
Purpose: An inverse planning simulated annealing (IPSA) algorithm for optimization of high-dose-rate (HDR) brachytherapy has been previously described. In this study, IPSA is compared with geometrical optimization (GO) for prostate brachytherapy.
Methods and materials: Using CT data collected from 10 patients, treatment plans were prepared using GO and IPSA. The clinical target volume (CTV) and critical organs (CO) including bladder, rectum, and urethra were contoured using Plato Version 14.2.1 (Nucletron Corp., Veenendaal, The Netherlands). Implant catheters were digitized using the CT planning system. All dwell positions outside of the CTV were turned off. Two optimized plans were generated for each implant using GO and IPSA. The same set of dose constraints were used for all inverse planning calculations and no manual adjustment of the dwell weight was used. Two prescription methods were used. Using the first method, coverage was prioritized: the prescription dose was normalized to the isodose volume that covers 98% of the CTV (V100 = 98% of CTV). The dose volume histograms (DVH) of CO were generated for comparison. Using the second method, sparing was prioritized: the prescription dose was normalized such that no urethra volume received 150% of the prescription dose (V150-urethra = 0 cc). The DVH of CTV and CO were generated, and the homogeneity index (HI) and conformal index (COIN) were calculated for comparison and compared using the Wilcoxon matched-pairs test.
Results: Using the coverage-prioritized method, the difference in V80-bladder dose was not statistically significant (p = 0.09; median: IPSA = 0.62 cc, GO = 1.05 cc). The V80-rectum ranged from 0.20-4.8 cc, and 0.05-1.4 cc using GO and IPSA, respectively. IPSA's V80-rectum was significantly lower (p = 0.005; median: IPSA=0.38 cc, GO = 1.31 cc). V150-urethra ranged from 0.02-0.75 cc and 0.0-0.01 cc using GO and IPSA, respectively. The V150-urethra was significantly lower using IPSA (p = 0.005; median: IPSA = 0.00 cc, GO = 0.33 cc). Using the sparing prioritized method, the V100-prostate ranged from 30-97% and 95-100% using GO and IPSA, respectively. This difference was statistically significant (p = 0.008). The HI and COIN were statistically higher using IPSA (p = 0.005).
Conclusion: Anatomy-based inverse optimization using IPSA is superior to dwell-position-based optimization using GO as it: (1) Improves target coverage and conformality while sparing normal structures, (2) Improves dose homogeneity within the target, and (3) Minimizes volume of non-contoured normal tissue irradiated. Routine application of three-dimensional brachytherapy planning and anatomy-based inverse dwell time optimization is recommended.
Similar articles
-
A comparison of anatomy-based inverse planning with simulated annealing and graphical optimization for high-dose-rate prostate brachytherapy.Brachytherapy. 2008 Jan-Mar;7(1):12-6. doi: 10.1016/j.brachy.2007.10.001. Epub 2007 Nov 26. Brachytherapy. 2008. PMID: 18037356
-
Anatomy-based inverse planning simulated annealing optimization in high-dose-rate prostate brachytherapy: significant dosimetric advantage over other optimization techniques.Int J Radiat Oncol Biol Phys. 2008 Nov 1;72(3):820-7. doi: 10.1016/j.ijrobp.2008.02.009. Epub 2008 May 1. Int J Radiat Oncol Biol Phys. 2008. PMID: 18455325
-
Evaluation of anatomy-based dwell position and inverse optimization in high-dose-rate brachytherapy of prostate cancer: a dosimetric comparison to a conventional cylindrical dwell position, geometric optimization, and dose-point optimization.Radiother Oncol. 2005 Jun;75(3):311-7. doi: 10.1016/j.radonc.2005.02.006. Radiother Oncol. 2005. PMID: 15890425
-
Pilot study in the treatment of endometrial carcinoma with 3D image-based high-dose-rate brachytherapy using modified Heyman packing: clinical experience and dose-volume histogram analysis.Int J Radiat Oncol Biol Phys. 2005 Jun 1;62(2):468-78. doi: 10.1016/j.ijrobp.2004.10.013. Int J Radiat Oncol Biol Phys. 2005. PMID: 15890589 Review.
-
[Physical brachytherapy planning].Praxis (Bern 1994). 2001 Sep 6;90(36):1532-43. Praxis (Bern 1994). 2001. PMID: 11593909 Review. German.
Cited by
-
Evaluation of hybrid inverse planning and optimization (HIPO) algorithm for optimization in real-time, high-dose-rate (HDR) brachytherapy for prostate.J Appl Clin Med Phys. 2013 Jul 8;14(4):4198. doi: 10.1120/jacmp.v14i4.4198. J Appl Clin Med Phys. 2013. PMID: 23835384 Free PMC article.
-
Dose error from deviation of dwell time and source position for high dose-rate 192Ir in remote afterloading system.J Radiat Res. 2014 Jul;55(4):780-7. doi: 10.1093/jrr/rru001. Epub 2014 Feb 23. J Radiat Res. 2014. PMID: 24566719 Free PMC article.
-
Dosimetry analyses comparing high-dose-rate brachytherapy, administered as monotherapy for localized prostate cancer, with stereotactic body radiation therapy simulated using CyberKnife.J Radiat Res. 2014 Nov;55(6):1114-21. doi: 10.1093/jrr/rru048. Epub 2014 Jun 23. J Radiat Res. 2014. PMID: 24957754 Free PMC article.
-
A fast multitarget inverse treatment planning strategy optimizing dosimetric measures for high-dose-rate (HDR) brachytherapy.Med Phys. 2017 Sep;44(9):4452-4462. doi: 10.1002/mp.12410. Epub 2017 Jul 20. Med Phys. 2017. PMID: 28626978 Free PMC article.
-
Comparison of manual and inverse optimisation techniques in high dose rate intracavitary brachytherapy of cervical cancer: A dosimetric study.Rep Pract Oncol Radiother. 2015 Sep-Oct;20(5):365-9. doi: 10.1016/j.rpor.2015.05.003. Epub 2015 Jun 26. Rep Pract Oncol Radiother. 2015. PMID: 26549994 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical