Safety and feasibility of using a central venous catheter for rapid contrast injection rates
- PMID: 15547237
- DOI: 10.2214/ajr.183.6.01831829
Safety and feasibility of using a central venous catheter for rapid contrast injection rates
Abstract
Objective: Our aim was to determine the safety and feasibility of using a central venous catheter for rapid contrast injections during CT.
Materials and methods: An in vitro experiment was performed using a 7-French Arrow-Howes multilumen central venous catheter. Each catheter port was tested by varying contrast agent flow rates delivered by a power injector. Contrast media specifications were kept similar to routine clinical practice. The in vivo experiment included 104 cases in which rapid contrast injections, 3.0-5.0 mL/sec, were delivered through a central venous catheter for dynamic CT examinations. Patient monitoring for early complications of contrast extravasation, cardiac arrhythmia, and allergic reactions was performed. Contrast injections were monitored for pressure limitation, automatic flow-rate adjustment, and catheter injury. Chart review was performed for delayed complications of mediastinal hematoma, infection, or catheter malfunction.
Results: During the in vitro experiment, all desired flow rates, 3.0-9.9 mL/sec, could be delivered through the central venous catheter with no catheter injury. No immediate or early patient or catheter complications were observed during the in vivo experiment. Follow-up evaluation revealed that 18 blood cultures and one catheter culture were positive for bacterial growth. In a subgroup of 43 patients, five contrast injections were pressure-limited by the power injector, and only one had the flow rate automatically adjusted to 3.6 mL/sec from 4.0 mL/sec.
Conclusion: Rapid contrast injection rates, at 3.0-5.0 mL/sec, through the Arrow-Howes multilumen central venous catheter are feasible and safe in the clinical setting. However, a strict protocol should be followed to avoid possible serious complications.
Similar articles
-
Power injection of contrast media using central venous catheters: feasibility, safety, and efficacy.AJR Am J Roentgenol. 2001 Feb;176(2):447-53. doi: 10.2214/ajr.176.2.1760447. AJR Am J Roentgenol. 2001. PMID: 11159092 Clinical Trial.
-
In-line pressures generated in small-bore central venous catheters during power injection of CT contrast media.Radiology. 1997 Jun;203(3):625-9. doi: 10.1148/radiology.203.3.9169679. Radiology. 1997. PMID: 9169679
-
Central venous catheter integrity during mechanical power injection of iodinated contrast medium.Radiology. 2009 Dec;253(3):870-8. doi: 10.1148/radiol.2533081086. Epub 2009 Sep 29. Radiology. 2009. PMID: 19789224
-
Risk of extravasation after power injection of contrast media via the proximal port of multilumen central venous catheters: case report and review of the literature.Rofo. 2010 Jan;182(1):14-9. doi: 10.1055/s-0028-1109742. Epub 2009 Oct 26. Rofo. 2010. PMID: 19859861 Review.
-
Rates of infection for single-lumen versus multilumen central venous catheters: a meta-analysis.Crit Care Med. 2003 Sep;31(9):2385-90. doi: 10.1097/01.CCM.0000084843.31852.01. Crit Care Med. 2003. PMID: 14501971 Review.
Cited by
-
The use of central venous catheters for intravenous contrast injection for CT examinations.Br J Radiol. 2011 Mar;84(999):197-203. doi: 10.1259/bjr/26062221. Br J Radiol. 2011. PMID: 21325362 Free PMC article. Review.
-
Plasticity of Candida albicans Biofilms.Microbiol Mol Biol Rev. 2016 Jun 1;80(3):565-95. doi: 10.1128/MMBR.00068-15. Print 2016 Sep. Microbiol Mol Biol Rev. 2016. PMID: 27250770 Free PMC article. Review.
-
Systematic review of the safety and efficacy of contrast injection via venous catheters for contrast-enhanced computed tomography.Eur J Radiol Open. 2017 Sep 29;4:118-122. doi: 10.1016/j.ejro.2017.09.002. eCollection 2017. Eur J Radiol Open. 2017. PMID: 29034281 Free PMC article. Review.
-
Effect of training and experience on qualitative and quantitative CT perfusion data.AJNR Am J Neuroradiol. 2007 Mar;28(3):428-32. AJNR Am J Neuroradiol. 2007. PMID: 17353307 Free PMC article.
-
Injection of gadolinium contrast through pediatric central venous catheters: a safety study.Pediatr Radiol. 2012 Sep;42(9):1064-9. doi: 10.1007/s00247-012-2397-z. Epub 2012 Apr 24. Pediatr Radiol. 2012. PMID: 22526282
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical