Impact factors on intraocular pressure measurements in healthy subjects
- PMID: 15548801
- PMCID: PMC1772426
- DOI: 10.1136/bjo.2004.049924
Impact factors on intraocular pressure measurements in healthy subjects
Abstract
Aim: To evaluate whether intraocular pressure (IOP) calculation by applanation tonometry is determined more essentially by the subject's neck position or by neck constriction.
Methods: 23 right eyes of 23 healthy subjects (12 male, 11 female) were included. IOP was measured by applanation tonometry with the TonoPen on sitting participants under four different conditions: with open collar upright (A) or with the head in the headrest of a slit lamp (B), with a tight necktie upright (C) or in slit lamp position (D). All measurements with neck constriction were performed 3 minutes after placing the necktie.
Results: Mean IOP was 16.9 (SD 2.3) mm Hg (range 11-21 mm Hg) (A), 18.1 (SD 2.2) mm Hg (range 14-22 mm Hg) (B), 17.9 (SD 2.9) mm Hg (range 12-25 mm Hg) (C) and 18.7 (SD 2.7) mm Hg (range 13-24 mm Hg) (D). Mean IOP increased by 1.3 (SD 2.6) mm Hg (p = 0.028, paired t test, range +0.2 to +2.4 mm Hg) if subjects changed position from A to B. There was no statistically significant difference between measurements with or without neck constriction.
Conclusion: Applanation tonometry may be inaccurate if performed in slit lamp position. In contrast, tight neckties do not significantly affect IOP evaluation in healthy subjects.
Comment in
-
Tight necktie, intraocular pressure, and intracranial pressure.Br J Ophthalmol. 2005 Jun;89(6):786-7; author reply 787. doi: 10.1136/bjo.2005.068544/10.1136/bjo.2005.069757. Br J Ophthalmol. 2005. PMID: 15923535 Free PMC article. No abstract available.
Similar articles
-
Effect of a tight necktie on intraocular pressure.Br J Ophthalmol. 2003 Aug;87(8):946-8. doi: 10.1136/bjo.87.8.946. Br J Ophthalmol. 2003. PMID: 12881330 Free PMC article.
-
[Comparison of precision of the TonoPenXL with the Goldmann and Draeger applanation tonometer in a sitting and recumbent position of the patients -- a clinical study on 251 eyes].Klin Monbl Augenheilkd. 2002 Nov;219(11):785-90. doi: 10.1055/s-2002-36317. Klin Monbl Augenheilkd. 2002. PMID: 12494368 German.
-
Effect of head position on intraocular pressure in horses.Am J Vet Res. 2006 Jul;67(7):1232-5. doi: 10.2460/ajvr.67.7.1232. Am J Vet Res. 2006. PMID: 16817748 Clinical Trial.
-
Clinical comparison of pascal dynamic contour tonometry and goldmann applanation tonometry in asymmetric open-angle glaucoma.J Glaucoma. 2007 Dec;16(8):694-9. doi: 10.1097/IJG.0b013e3180408dc6. J Glaucoma. 2007. PMID: 18091457
-
Comparison of Goldmann applanation and applanation resonance tonometry in a biomicroscope-based in vitro porcine eye model.J Med Eng Technol. 2006 Nov-Dec;30(6):345-52. doi: 10.1080/03091900512331332528. J Med Eng Technol. 2006. PMID: 17060162
Cited by
-
Tight necktie, intraocular pressure, and intracranial pressure.Br J Ophthalmol. 2005 Jun;89(6):786-7; author reply 787. doi: 10.1136/bjo.2005.068544/10.1136/bjo.2005.069757. Br J Ophthalmol. 2005. PMID: 15923535 Free PMC article. No abstract available.
-
Comparison of three methods of tonometry in normal subjects: Goldmann applanation tonometer, non-contact airpuff tonometer, and Tono-Pen XL.Clin Ophthalmol. 2014 Jun 7;8:1069-74. doi: 10.2147/opth.s6391. eCollection 2014. Clin Ophthalmol. 2014. PMID: 24944507 Free PMC article.
-
Noninvasive intracranial pressure monitoring via optic nerve sheath diameter for robotic surgery in steep Trendelenburg position.Saudi J Anaesth. 2015 Jul-Sep;9(3):239-46. doi: 10.4103/1658-354X.154693. Saudi J Anaesth. 2015. PMID: 26240539 Free PMC article.
References
-
- American Academy of Ophthalmology. Intraocular pressure and aqueous humor dynamics. In: Liesegang TJ, Deutsch TA, Grand MG, eds. Basic and clinical science course. Section 10: Glaucoma. San Francisco: American Academy of Ophthalmology, 2003:14–24.
-
- Emara BY, Tingey DP, Probst LE, et al. Central corneal thickness in low-tension glaucoma. Can J Ophthalmol 1999;34:319–24. - PubMed
-
- Viestenz A , Lausen B, Junemann AM, et al. Comparison of precision of the TonoPenXL with the Goldmann and Draeger applanation tonometer in a sitting and recumbent position of the patients—a clinical study on 251 eyes. Klin Monatsbl Augenheilkd 2002;219:785–90. - PubMed
-
- Sultan M , Blondeau P. Episcleral venous pressure in younger and older subjects in the sitting and supine positions. J Glaucoma 2003;12:370–3. - PubMed
MeSH terms