4 T actively detunable transmit/receive transverse electromagnetic coil and 4-channel receive-only phased array for (1)H human brain studies
- PMID: 15562466
- DOI: 10.1002/mrm.20264
4 T actively detunable transmit/receive transverse electromagnetic coil and 4-channel receive-only phased array for (1)H human brain studies
Abstract
The design and construction of a 4 T transverse electromagnetic (TEM) transmit/receive head coil and a four-channel phased array receive-only RF system are described. To enable both high-resolution imaging of the entire brain and high-resolution spectroscopic imaging, active PIN diode decoupling was used in both the TEM resonator and each surface coil in the array. This configuration allows for both transmission and reception from the volume coil as well as reception from the phased array. The surface coils were decoupled by overlapping the coils and using preamplifier decoupling. Since at high frequencies construction of a lumped element matching quarter wavelength transformer, an important component of the preamplifier decoupling, becomes difficult, a transmission line approach was used. The system was tested and compared to a TEM volume transmit/receive head coil. A four- to sixfold improvement in signal-to-noise ratio from the sensitive volume of the array was achieved.
(c) 2004 Wiley-Liss, Inc.
Similar articles
-
High-field head radiofrequency volume coils using transverse electromagnetic (TEM) and phased array technologies.NMR Biomed. 2009 Nov;22(9):960-74. doi: 10.1002/nbm.1262. NMR Biomed. 2009. PMID: 18574792
-
Eight-channel phased array coil and detunable TEM volume coil for 7 T brain imaging.Magn Reson Med. 2005 Jul;54(1):235-40. doi: 10.1002/mrm.20547. Magn Reson Med. 2005. PMID: 15968650 Clinical Trial.
-
SENSE imaging with a quadrature half-volume transverse electromagnetic (TEM) coil at 4T.J Magn Reson Imaging. 2006 Oct;24(4):934-8. doi: 10.1002/jmri.20714. J Magn Reson Imaging. 2006. PMID: 16958065
-
Theory and application of array coils in MR spectroscopy.NMR Biomed. 1997 Dec;10(8):394-410. doi: 10.1002/(sici)1099-1492(199712)10:8<394::aid-nbm494>3.0.co;2-0. NMR Biomed. 1997. PMID: 9542737 Review.
-
Artifacts by Misalignment of Cardiac Magnetic Resonance Phased-array Coil Elements: From Simulation to In vivo Test.Curr Med Imaging Rev. 2019;15(3):301-307. doi: 10.2174/1573405613666171024150250. Curr Med Imaging Rev. 2019. PMID: 31989881 Review.
Cited by
-
Occipital levels of GABA are related to severe headaches in migraine.Neurology. 2008 May 27;70(22):2078-80. doi: 10.1212/01.wnl.0000313376.07248.28. Neurology. 2008. PMID: 18505983 Free PMC article. No abstract available.
-
Comparison of three commercially available radio frequency coils for human brain imaging at 3 Tesla.MAGMA. 2008 Mar;21(1-2):53-61. doi: 10.1007/s10334-007-0100-4. Epub 2008 Jan 10. MAGMA. 2008. PMID: 18193311
-
Improved homogeneity of the transmit field by simultaneous transmission with phased array and volume coil.J Magn Reson Imaging. 2010 Aug;32(2):476-81. doi: 10.1002/jmri.22257. J Magn Reson Imaging. 2010. PMID: 20677280 Free PMC article.
-
Actively Decoupled Transmit-Receive Coil-Pair for Mouse Brain MRI.Concepts Magn Reson Part B Magn Reson Eng. 2008 Oct;33B(4):252-259. doi: 10.1002/cmr.b.20124. Concepts Magn Reson Part B Magn Reson Eng. 2008. PMID: 26207105 Free PMC article.
-
Multiple-Input Multiple-Output (MIMO) MRI: Combining Parallel Excitation and Parallel Reception for Enhanced Imaging.IEEE Trans Comput Imaging. 2019 Dec;5(4):596-605. doi: 10.1109/tci.2019.2904882. Epub 2019 Mar 13. IEEE Trans Comput Imaging. 2019. PMID: 31875167 Free PMC article.
Publication types
MeSH terms
Grants and funding
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical