[Quality evaluation of three different sperm counting chambers]
- PMID: 15562789
[Quality evaluation of three different sperm counting chambers]
Abstract
Objective: Semen evaluation is the most important laboratory test for assessing male fertility. However, lack of strict quality control (QC) for semen analyses in hospital andrology laboratories makes it difficult and meaningless to compare semen data between different laboratories. This paper reports a comparative study on the accuracy of the Hemacytometer (Qiujing Inc., Shanghai, China), Makler (Sefi-Medical Instrument, Haifa, Israel), and Cell-VU (Millennium Sciences Inc., New York, USA) chambers for sperm counting.
Methods: Both low [(18 +/- 2.5) x 10(6)/ml] and high [(35 +/- 5) x 10(6)/ml] pre-calibrated standard latex bead solutions (Hamilton Thorne Biosciences, USA) were used as the quality control solution to perform counts on the three different counting chambers. Bead counts for the three different chambers were compared, and variability within the chambers determined for standard solutions at low and high concentrations of latex beads, respectively.
Results: Mean bead concentrations for the Cell-VU, Hemacytometer and Makler chambers were (37.63 +/- 4.89), (42.74 +/- 4.98) and (53.52 +/- 6.67) x 10(6)/ml respectively for a standard solution containing (35 +/- 5) x 10(6) beads/ml, and (18.22 +/- 1.77), (20.48 +/- 1.56), (24.97 +/- 4.75) x 10(6)/ml respectively for a standard solution containing (18 +/- 2.5) x 10(6) beads/ml. Mean bead concentrations for the Cell-VU chamber were consistently similar and close to the standard pre-calibrated bead solutions, while those for both the Hemacytometer and the Makler chambers were significantly overestimated (P < 0.001). The average coefficients of variation for the Cell-VU chamber were 7.51% for a higher concentration of the standard solution containing (36 +/- 5) x 10(6) beads/ml and 1.22% for a lower concentration of the standard solution containing (18 +/- 2.5) x 10(6) beads/ml, while the mean variation rates of the Hemacytometer and Makler chambers were 22.11% and 13.78% for a standard solution containing (36 +/- 5) x 10(6) beads/ml, and 52.91% and 38.72% for a standard solution containing (18 +/- 2.5) x 10(6) beads/ ml, respectively.
Conclusion: Semen analysis is one of the most important tests for male fertility evaluation, but the data obtained from commercially available counting chambers may differ markedly in accuracy and reliability. Results from this comparative study demonstrated that the Cell-VU chamber exhibits significantly more accurate and less variable results than those of the Hemacytometer and Makler chambers. To ensure the best possible evaluations and accurate diagnoses, we therefore recommend that Cell-VU be used as the standard counting chamber for routine semen analyses in andrology laboratories.
Similar articles
-
[Comparison of four methods for sperm counting].Zhonghua Nan Ke Xue. 2006 Mar;12(3):222-4, 227. Zhonghua Nan Ke Xue. 2006. PMID: 16597036 Chinese.
-
Accuracy of semen counting chambers as determined by the use of latex beads.Fertil Steril. 1996 Oct;66(4):662-5. doi: 10.1016/s0015-0282(16)58587-2. Fertil Steril. 1996. PMID: 8816636
-
Is flow cytometry really adapted to the determination of sperm concentration?Scand J Clin Lab Invest. 2007;67(4):394-401. doi: 10.1080/00365510601124032. Scand J Clin Lab Invest. 2007. PMID: 17558894
-
[Advances in standardization and quality control for the analysis of sperm quality parameters].Zhonghua Nan Ke Xue. 2007 Nov;13(11):963-8. Zhonghua Nan Ke Xue. 2007. PMID: 18077904 Review. Chinese.
-
Can external quality control improve pig AI efficiency?Theriogenology. 2008 Nov;70(8):1346-51. doi: 10.1016/j.theriogenology.2008.06.006. Epub 2008 Jul 24. Theriogenology. 2008. PMID: 18656253 Review.
Cited by
-
A survey on the status of semen analysis in 118 laboratories in China.Asian J Androl. 2010 Jan;12(1):104-10. doi: 10.1038/aja.2008.41. Epub 2009 Feb 23. Asian J Androl. 2010. PMID: 19234484 Free PMC article.
-
Accuracy Evaluation of The Depth of Six Kinds of Sperm Counting Chambers for both Manual and Computer-Aided Semen Analyses.Int J Fertil Steril. 2016 Jan-Mar;9(4):527-33. doi: 10.22074/ijfs.2015.4612. Epub 2015 Dec 23. Int J Fertil Steril. 2016. PMID: 26985341 Free PMC article.