. 2004 Nov 29;4(1):8.
doi: 10.1186/1471-2253-4-8.

Endotracheal tube cuff pressure in three hospitals, and the volume required to produce an appropriate cuff pressure

Affiliations

Endotracheal tube cuff pressure in three hospitals, and the volume required to produce an appropriate cuff pressure

Papiya Sengupta et al. BMC Anesthesiol. .

Abstract

BACKGROUND: Cuff pressure in endotracheal (ET) tubes should be in the range of 20-30 cm H2O. We tested the hypothesis that the tube cuff is inadequately inflated when manometers are not used. METHODS: With IRB approval, we studied 93 patients under general anesthesia with an ET tube in place in one teaching and two private hospitals. Anesthetists were blinded to study purpose. Cuff pressure in tube sizes 7.0 to 8.5 mm was evaluated 60 min after induction of general anesthesia using a manometer connected to the cuff pilot balloon. Nitrous oxide was disallowed. After deflating the cuff, we reinflated it in 0.5-ml increments until pressure was 20 cmH2O. RESULTS: Neither patient morphometrics, institution, experience of anesthesia provider, nor tube size influenced measured cuff pressure (35.3 +/- 21.6 cmH2O). Only 27% of pressures were within 20-30 cmH2O; 27% exceeded 40 cmH2O. Although it varied considerably, the amount of air required to achieve a cuff pressure of 20 cmH2O was similar with each tube size. CONCLUSION: We recommend that ET cuff pressure be set and monitored with a manometer.

PubMed Disclaimer

Figures

Figure 1
Figure 1
The relationship between measured cuff pressure and volume of air in the cuff. There was a linear relationship between measured cuff pressure (cmH2O) and volume (ml) of air removed from the cuff: Pressure = 7.5. Volume + 2.7, r2 = 0.39.

References

    1. Tobin MJ, Grenvik A. Nosocomial lung infection and its diagnosis. Crit Care Med. 1984;12:191–199. - PubMed
    1. Dullenkopf A, Gerber A, Weiss M. Fluid leakage past tracheal tube cuffs: evaluation of the new Microcuff endotracheal tube. Intensive Care Med. 2003;29:1849–1853. doi: 10.1007/s00134-003-1933-6. - DOI - PubMed
    1. Lien TC, Wang JH. [Incidence of pulmonary aspiration with different kinds of artificial airways] Zhonghua Yi Xue Za Zhi (Taipei) 1992;49:348–353. - PubMed
    1. Gottschalk A, Burmeister MA, Blanc I, Schulz F, Standl T. [Rupture of the trachea after emergency endotracheal intubation] Anasthesiol Intensivmed Notfallmed Schmerzther. 2003;38:59–61. doi: 10.1055/s-2003-36557. - DOI - PubMed
    1. Striebel HW, Pinkwart LU, Karavias T. [Tracheal rupture caused by overinflation of endotracheal tube cuff] Anaesthesist. 1995;44:186–188. doi: 10.1007/s001010050146. - DOI - PubMed

LinkOut - more resources