Non-invasive assessment of mitral valve area during percutaneous balloon mitral valvuloplasty: role of real-time 3D echocardiography
- PMID: 15571823
- DOI: 10.1016/j.ehj.2004.09.041
Non-invasive assessment of mitral valve area during percutaneous balloon mitral valvuloplasty: role of real-time 3D echocardiography
Abstract
Background: In the last decade, multiple studies depicted discrepancies between mitral valvular orifice area (MVA) measurements obtained with the pressure half-time (PHT) method and invasive methods during the immediate post-percutaneous mitral valvuloplasty (PMV) period. Our aim was to assess the accuracy of Real-Time 3D echo (RT3D) to measure the MVA in the immediate post-PMV period. The invasively determined MVA was used as the gold standard.
Methods and results: We studied 29 patients with rheumatic mitral stenosis from two centres (27 women; mean age 48.2+/-11.3 years), all of which had underwent PMV. MVA was calculated before and after PMV using the PHT method, 2D echo planimetry, RT3D echo planimetry and invasive determination (Gorlin's method). The RT3D MVA assessment showed a better agreement with the invasively derived MVA before and in the immediate post-PMV period (Bland-Altman analysis: Average difference between both methods and limits of agreement: 0.01 (-0.31 to 0.33) cm(2) and -0.12 (-0.71 to 0.47) cm(2)) before and immediately after the PMV, respectively.
Conclusions: RT3D is a feasible and accurate technique for measuring MVA in patients with RMVS. It has the best agreement with the invasively determined MVA, particularly in the immediate post-PMV period.
Comment in
-
Should mitral valve area assessment in patients with mitral stenosis be based on anatomical or on functional evaluation? A plea for 3D echocardiography as the new clinical standard.Eur Heart J. 2004 Dec;25(23):2073-4. doi: 10.1016/j.ehj.2004.10.001. Eur Heart J. 2004. PMID: 15571820 No abstract available.
Similar articles
-
Real-time three-dimensional echocardiography for rheumatic mitral valve stenosis evaluation: an accurate and novel approach.J Am Coll Cardiol. 2004 Jun 2;43(11):2091-6. doi: 10.1016/j.jacc.2004.01.046. J Am Coll Cardiol. 2004. PMID: 15172418
-
Real-time 3D transesophageal echocardiography for the evaluation of rheumatic mitral stenosis.JACC Cardiovasc Imaging. 2011 Jun;4(6):580-8. doi: 10.1016/j.jcmg.2010.12.009. JACC Cardiovasc Imaging. 2011. PMID: 21679891
-
A novel method to measure mitral valve area in patients with rheumatic mitral stenosis using three-dimensional transesophageal echocardiography: Feasibility and validation.Echocardiography. 2018 Mar;35(3):368-374. doi: 10.1111/echo.13786. Epub 2017 Dec 22. Echocardiography. 2018. PMID: 29271495
-
Three-dimensional echocardiography for assessment of mitral valve stenosis.Curr Opin Cardiol. 2009 Sep;24(5):415-9. doi: 10.1097/HCO.0b013e32832e165b. Curr Opin Cardiol. 2009. PMID: 19550310 Review.
-
The use of three-dimensional echocardiography for the evaluation of and treatment of mitral stenosis.Cardiol Clin. 2007 May;25(2):311-8. doi: 10.1016/j.ccl.2007.06.008. Cardiol Clin. 2007. PMID: 17765111 Review.
Cited by
-
Real-time three-dimensional transthoracic echocardiography in daily practice: initial experience.Cardiovasc Ultrasound. 2012 Mar 26;10:14. doi: 10.1186/1476-7120-10-14. Cardiovasc Ultrasound. 2012. PMID: 22448671 Free PMC article.
-
Editorial: Percutaneous mitral valvotomy: Balloon made in Japan still dominates the world.J Cardiol Cases. 2016 Apr 8;13(6):169-170. doi: 10.1016/j.jccase.2016.03.009. eCollection 2016 Jun. J Cardiol Cases. 2016. PMID: 30546637 Free PMC article. No abstract available.
-
Assessment of aortic stenosis by three-dimensional echocardiography: an accurate and novel approach.Heart. 2007 Jul;93(7):801-7. doi: 10.1136/hrt.2006.110726. Epub 2007 May 8. Heart. 2007. PMID: 17488766 Free PMC article.
-
Role of modern 3D echocardiography in valvular heart disease.Korean J Intern Med. 2014 Nov;29(6):685-702. doi: 10.3904/kjim.2014.29.6.685. Epub 2014 Oct 31. Korean J Intern Med. 2014. PMID: 25378966 Free PMC article. Review.
-
Procedural guidance using advance imaging techniques for percutaneous edge-to-edge mitral valve repair.Curr Cardiol Rep. 2014 Feb;16(2):452. doi: 10.1007/s11886-013-0452-5. Curr Cardiol Rep. 2014. PMID: 24430014 Review.