Measurement of cardiac output before and after cardiopulmonary bypass: Comparison among aortic transit-time ultrasound, thermodilution, and noninvasive partial CO2 rebreathing
- PMID: 15578466
- DOI: 10.1053/j.jvca.2004.07.005
Measurement of cardiac output before and after cardiopulmonary bypass: Comparison among aortic transit-time ultrasound, thermodilution, and noninvasive partial CO2 rebreathing
Abstract
Objectives: A noninvasive continuous cardiac output system (NICO) has been developed recently. NICO uses a ratio of the change in the end-tidal carbon dioxide partial pressure and carbon dioxide elimination in response to a brief period of partial rebreathing to measure CO. The aim of this study was to compare the agreement among NICO, bolus (TDCO), and continuous thermodilution (CCO), with transit-time flowmetry of the ascending aorta using an ultrasonic flow probe (UFP) before and after cardiopulmonary bypass (CPB).
Design: Prospective, observational human study.
Setting: Veterans Affairs Medical Center Hospital.
Participants: Sixty-eight patients.
Methods: Matched sets of CO measurements between NICO, TDCO, CCO, and UFP were collected in 68 patients undergoing elective CABG at specific time periods before and after separation from CPB. After anesthetic induction, all patients had an NICO sensor attached between the endotracheal tube and the breathing circuit, a PAC floated into the pulmonary artery for TDCO and CCO monitoring, and a UFP positioned on the ascending aorta and used for the reference CO. Bland-Altman analysis was used to compare the agreement among the different methods.
Measurements and main results: Bland-Altman analysis of CO measurements before CPB yielded a bias, precision, and percent error of 0.04 L/min +/- 1.07 L/min (44.8%) for NICO, 0.18 L/min +/- 1.01 L/min (41.7%) for TDCO, and 0.29 L/min +/- 1.40 L/min (57.5%) for CCO compared with simultaneous UFP CO measurements, respectively. After separation from CPB (average 29 mins), bias, precision, and percent error were -0.46 L/min +/- 1.06 L/min (37.3%) for NICO, 0.35 L/min +/- 1.39 L/min (46.1%) for TDCO, and 0.36 L/min +/- 1.96 L/min (64.7%) for CCO compared with UFP CO measurements, respectively.
Conclusions: Before initiation of CPB, the accuracy for all 3 techniques was similar. After separation from CPB, the tendency was for NICO to underestimate CO and for TDCO and CCO to overestimate it. NICO offers an alternative to invasive CO measurement.
Similar articles
-
New generation continuous cardiac output monitoring from carbon dioxide elimination.BMC Anesthesiol. 2019 Feb 26;19(1):28. doi: 10.1186/s12871-019-0699-5. BMC Anesthesiol. 2019. PMID: 30808309 Free PMC article. Review.
-
Partial CO2 rebreathing indirect Fick technique for non-invasive measurement of cardiac output.J Clin Monit Comput. 2000;16(5-6):361-74. doi: 10.1023/a:1011403717822. J Clin Monit Comput. 2000. PMID: 12580219
-
Performance of noninvasive partial CO2 rebreathing cardiac output and continuous thermodilution cardiac output in patients undergoing aortic reconstruction surgery.Anesthesiology. 2003 Aug;99(2):283-8. doi: 10.1097/00000542-200308000-00009. Anesthesiology. 2003. PMID: 12883400 Clinical Trial.
-
Evaluation of partial carbon dioxide rebreathing cardiac output measurement during thoracic surgery.J Cardiothorac Vasc Anesth. 2007 Oct;21(5):655-8. doi: 10.1053/j.jvca.2007.01.012. Epub 2007 Apr 11. J Cardiothorac Vasc Anesth. 2007. PMID: 17905269
-
Noninvasive Cardiac Output Monitoring in Cardiothoracic Surgery Patients: Available Methods and Future Directions.J Cardiothorac Vasc Anesth. 2019 Jun;33(6):1742-1752. doi: 10.1053/j.jvca.2018.06.012. Epub 2018 Jun 27. J Cardiothorac Vasc Anesth. 2019. PMID: 30318422 Review.
Cited by
-
The contemporary pulmonary artery catheter. Part 2: measurements, limitations, and clinical applications.J Clin Monit Comput. 2022 Feb;36(1):17-31. doi: 10.1007/s10877-021-00673-5. Epub 2021 Mar 1. J Clin Monit Comput. 2022. PMID: 33646499 Free PMC article. Review.
-
Cardiac Output Measurements in Septic Patients: Comparing the Accuracy of USCOM to PiCCO.Crit Care Res Pract. 2012;2012:270631. doi: 10.1155/2012/270631. Epub 2011 Nov 29. Crit Care Res Pract. 2012. PMID: 22191019 Free PMC article.
-
New generation continuous cardiac output monitoring from carbon dioxide elimination.BMC Anesthesiol. 2019 Feb 26;19(1):28. doi: 10.1186/s12871-019-0699-5. BMC Anesthesiol. 2019. PMID: 30808309 Free PMC article. Review.
-
Continuous minimally invasive peri-operative monitoring of cardiac output by pulmonary capnotracking: comparison with thermodilution and transesophageal echocardiography.J Clin Monit Comput. 2012 Apr;26(2):121-32. doi: 10.1007/s10877-012-9342-4. Epub 2012 Feb 18. J Clin Monit Comput. 2012. PMID: 22350312
-
Evaluation of a CO2 partial rebreathing functional residual capacity measurement method for use during mechanical ventilation.J Clin Monit Comput. 2011 Dec;25(6):397-404. doi: 10.1007/s10877-011-9318-9. Epub 2011 Nov 6. J Clin Monit Comput. 2011. PMID: 22057246