Comparative Study
. 2004 Dec;42(12):5489-92.
doi: 10.1128/JCM.42.12.5489-5492.2004.

Evaluation of a bedside immunochromatographic test for detection of adenovirus in respiratory samples, by comparison to virus isolation, PCR, and real-time PCR

Affiliations
Comparative Study

Evaluation of a bedside immunochromatographic test for detection of adenovirus in respiratory samples, by comparison to virus isolation, PCR, and real-time PCR

Tsuguto Fujimoto et al. J Clin Microbiol. 2004 Dec.

Abstract

An immunochromatography (IC) kit for human adenovirus (HAdV) was evaluated with 138 patient nasopharyngeal samples. The samples were collected at a sentinel clinic in Japan from January through June 2003. Patients were diagnosed by clinical manifestation of pharyngoconjunctival fever (n = 38) or exudative tonsillitis (n = 100). The IC kit was positive for 84% (116 of 138) of patients diagnosed at bedside. The remaining extract solution of the IC kit test was transferred into maintenance medium and tested via laboratory diagnoses. The IC kit had 95% sensitivity (116 of 122 patients) with HAdV isolation (isolation) as the standard and 91% sensitivity (116 of 128 patients) with PCR as the standard. All of the IC kit-positive samples were isolation and PCR positive. Similarly, all the isolation-positive samples were PCR positive. Twenty-two IC kit-negative samples were evaluated by real-time PCR. Six samples were IC kit negative and isolation positive and contained 3.8 x 10(7) to 2.5 x 10(9) copies of the HAdV genome/ml. Five samples that were only PCR positive contained 3.0 x 10(4) to 3.8 x 10(5) copies of the HAdV genome/ml, but one sample was real-time PCR negative. We conclude that the IC kit is a useful bedside diagnostic tool for HAdV infections because it has 95% sensitivity (compared to isolation), but a negative result does not always rule out HAdV infection.

PubMed Disclaimer

Figures

FIG. 1.
FIG. 1.
Detection of adenovirus in a clinical sample by the IC kit. The test is positive if two colored lines appear in the sample (S) and control (C) areas. The test is negative when only one colored line appears in the control area.

References

    1. Echavarria, M., M. Forman, J. Ticehursst, J. S. Dulmer, and P. Charache. 1998. PCR method for detection of adenovirus in urine of healthy and human immunodeficiency virus-infected individuals, J. Clin. Microbiol. 36:3323-3326. - PMC - PubMed
    1. Echavarria, M., M. Forman, M. J. van Tol, J. M. Vossen, P. Charache, and A. C. Kroes. 2001. Prediction of severe disseminated adenovirus infection by serum PCR. Lancet 358:384-385. - PubMed
    1. Fujimoto, T., M. Chikahira, T. Kase, S. Morikawa, T. Okafuji, Y. Yokota, and O. Nishio. 2000. Single-tube multiplex PCR for rapid and sensitive diagnosis of subgenus B and other subgenera adenoviruses in clinical samples. Microbiol. Immunol. 44:821-826. - PubMed
    1. Fujimoto, T., M. Chikahira, S. Yoshida, H. Ebira, A. Hasegawa, A. Totsuka, and O. Nishio. 2002. Outbreak of central nervous system disease associated with hand, foot, and mouth disease in Japan during the summer of 2000: detection and molecular epidemiology of enterovirus 71. Microbiol. Immunol. 46:621-627. - PubMed
    1. Gu, Z., S. W. Belzer, C. S. Gibson, M. J. Bankowski, and R. T. Hayden. 2003. Multiplexed, real-time PCR for quantitative detection of human adenovirus. J. Clin. Microbiol. 41:4636-4641. - PMC - PubMed

Publication types

MeSH terms

Substances

LinkOut - more resources