Abstract
Objectives: To evaluate the impact on photostimulable phosphor (PSP) image signal-to-noise ratio (SNR) of pre-scanning ambient lighting exposures.
Method: PSP imaging plates (IPs) were exposed to different radiation exposures to achieve flat field images. The exposed IPs were subjected variously to visible light of different intensities (300, 150 or 20 lux) for durations ranging from < 10 s to 120 s. They were processed using laser scanners from two systems for further comparison (DenOptix versus ScanX). Histogram analysis was performed in each case and mean pixel value and its standard deviation were used as surrogates to assess SNR. Statistical methods applied included analysis of variance with Tukey honestly significant difference test for pair wise comparisons. The a priori alpha was set at P < or = 0.05.
Results: SNR decreased with increased duration and intensity of pre-scanning light exposure. Lower X-ray exposures resulted in decreased signal resulting in reduced SNR, and increased the need to reduce ambient lighting. No statistically significant differences were found comparing ScanX and DenOptix digital imaging systems in terms of SNR.
Conclusion: Reduced ambient lighting is preferred for handling IPs prior to processing in the laser scanner.
Similar articles
-
Time and motion study: a comparison of two photostimulable phosphor imaging systems used in dentistry.Dentomaxillofac Radiol. 2006 Sep;35(5):315-8. doi: 10.1259/dmfr/29518441. Dentomaxillofac Radiol. 2006. PMID: 16940478
-
Performance evaluation and testing of digital intra-oral radiographic systems.Radiat Prot Dosimetry. 2005;117(1-3):313-7. doi: 10.1093/rpd/nci762. Epub 2006 Feb 3. Radiat Prot Dosimetry. 2005. PMID: 16461488
-
How Does Ambient Light Affect the Image Quality of Phosphor Plate Digital Radiography? A Quantitative Analysis Using Contemporary Digital Radiographic Systems.Sensors (Basel). 2022 Nov 9;22(22):8627. doi: 10.3390/s22228627. Sensors (Basel). 2022. PMID: 36433224 Free PMC article.
-
Comments on noise and resolution of the DenOptix radiography system.Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2003 Jun;95(6):746-51. doi: 10.1067/moe.2003.245. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2003. PMID: 12789159
-
Exposure indicator degradation from CR plate processing delays.Radiol Technol. 2009 May-Jun;80(5):401-9. Radiol Technol. 2009. PMID: 19457844 Review.
Cited by
-
Effect of alternative photostimulable phosphor plates erasing times on subjective digital image quality.Dentomaxillofac Radiol. 2010 Jan;39(1):23-7. doi: 10.1259/dmfr/49065239. Dentomaxillofac Radiol. 2010. PMID: 20089740 Free PMC article.
-
The effect of delayed scanning of storage phosphor plates on occlusal caries detection.Dentomaxillofac Radiol. 2012 May;41(4):309-15. doi: 10.1259/dmfr/12935491. Epub 2012 Jan 26. Dentomaxillofac Radiol. 2012. PMID: 22282506 Free PMC article.
-
Enterprise-wide implementation of digital radiography in oral and maxillofacial imaging: the University of Florida Dentistry System.J Digit Imaging. 2009 Jun;22(3):232-41. doi: 10.1007/s10278-008-9149-5. Epub 2008 Sep 3. J Digit Imaging. 2009. PMID: 18769969 Free PMC article.
-
Intraoral digital radiography: A comprehensive report on the technical specifications of current and historical systems.Imaging Sci Dent. 2025 Mar;55(1):72-89. doi: 10.5624/isd.20240248. Epub 2025 Mar 10. Imaging Sci Dent. 2025. PMID: 40191390 Free PMC article.
-
Can Digital Enhancement Restore the Image Quality of Phosphor Plate-Based Radiographs Partially Damaged by Ambient Light?J Imaging Inform Med. 2024 Feb;37(1):145-150. doi: 10.1007/s10278-023-00922-w. Epub 2024 Jan 12. J Imaging Inform Med. 2024. PMID: 38343236 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Miscellaneous