Effects of ground loop currents on signal intensities in electrospray mass spectrometry
- PMID: 15589752
- DOI: 10.1016/j.jasms.2004.08.005
Effects of ground loop currents on signal intensities in electrospray mass spectrometry
Abstract
The occurrence of electrochemical processes during the operation of an electrospray ionization (ESI) source is well established. In the positive ion mode, electrons are drawn from the ESI metal capillary to a high voltage power supply. These electrons are the product of charge-balancing oxidation reactions taking place at the liquid/metal interface of the ion source. In a recent study, (Anal. Chem.2001, 73, 4836-4844), our group has shown that the introduction of a ground loop can dramatically enhance the rate of these oxidation processes. Such a ground loop can be introduced by connecting the sample infusion syringe (or the liquid chromatography column, in the case of LC-MS studies) to ground. The magnitude of the ground loop current can be controlled by the electrolyte concentration in the analyte solution, and by the dimensions of the capillary connecting the syringe needle and the ESI source. Using ferrocene as a model system, it is demonstrated that the introduction of such a ground loop can significantly enhance the signal intensity of analytes that form electrochemically ionized species during ESI. However, analytes that form protonated molecular ions, such as reserpine, also show higher signal intensities when a ground loop is introduced into the system. This latter observation is attributed to the occurrence of electrolytic solvent (acetonitrile and/or water) oxidation processes. These reactions generate protons within the ion source, and thus facilitate the formation of [M + nH](n+) ions. Overall, this work provides an example of how the careful control of electrochemical parameters can be exploited to optimize signal intensities in ESI-MS.
Similar articles
-
Mass spectrometric methods for monitoring redox processes in electrochemical cells.Mass Spectrom Rev. 2015 Jan-Feb;34(1):64-92. doi: 10.1002/mas.21409. Epub 2013 Dec 10. Mass Spectrom Rev. 2015. PMID: 24338642 Free PMC article. Review.
-
Electrochemically induced pH changes resulting in protein unfolding in the ion source of an electrospray mass spectrometer.Anal Chem. 2001 Oct 15;73(20):4836-44. doi: 10.1021/ac010545r. Anal Chem. 2001. PMID: 11681459
-
Simultaneous detection of polar and nonpolar compounds by ambient mass spectrometry with a dual electrospray and atmospheric pressure chemical ionization source.Anal Chem. 2015 Feb 3;87(3):1743-8. doi: 10.1021/ac503625m. Epub 2015 Jan 20. Anal Chem. 2015. PMID: 25562530
-
Efficient analyte oxidation in an electrospray ion source using a porous flow-through electrode emitter.J Am Soc Mass Spectrom. 2004 Dec;15(12):1755-66. doi: 10.1016/j.jasms.2004.08.013. J Am Soc Mass Spectrom. 2004. PMID: 15589753
-
A review of recent trends in electrospray ionisation-mass spectrometry for the analysis of metal-organic ligand complexes.Anal Chim Acta. 2010 Sep 30;678(2):140-8. doi: 10.1016/j.aca.2010.08.023. Epub 2010 Aug 26. Anal Chim Acta. 2010. PMID: 20888445 Review.
Cited by
-
2-Nitrophenyl Aryl Sulfides Undergo Both Intramolecular and Electrospray-Induced Intermolecular Oxidation of Sulfur: An Experimental and Theoretical Case Study.Int J Mass Spectrom. 2009 Jun 1;283(1-3):222-228. doi: 10.1016/j.ijms.2009.04.003. Int J Mass Spectrom. 2009. PMID: 20161084 Free PMC article.
-
Mass spectrometric methods for monitoring redox processes in electrochemical cells.Mass Spectrom Rev. 2015 Jan-Feb;34(1):64-92. doi: 10.1002/mas.21409. Epub 2013 Dec 10. Mass Spectrom Rev. 2015. PMID: 24338642 Free PMC article. Review.
References
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources