Use of the Continuous Glucose Monitoring System to guide therapy in patients with insulin-treated diabetes: a randomized controlled trial
- PMID: 15595336
- DOI: 10.4065/79.12.1521
Use of the Continuous Glucose Monitoring System to guide therapy in patients with insulin-treated diabetes: a randomized controlled trial
Abstract
Objective: To show improved glycemic control in patients with insulin-treated diabetes after adjustments to the diabetes management plan based on either continuous glucose monitoring using the Continuous Glucose Monitoring System (CGMS) or frequent self-monitoring of blood glucose (SMBG) using a home blood glucose meter.
Patients and methods: From January to September 2000, patients aged 19 to 76 years with insulin-treated diabetes were assigned to insulin therapy adjustments based on either CGMS or SMBG values. At the end of the study, patients in both groups used the CGMS for 3 days; these values were used to calculate measures of hypoglycemia. Repeated-measures analysis of variance with post hoc comparisons were used to test differences in hemoglobin A1c levels and hypoglycemia between the 2 study groups.
Results: A total of 128 patients were enrolled in the study. Nineteen discontinued study participation, leaving 51 in the CGMS group and 58 in the SMBG group. No significant differences were noted in demographics or baseline characteristics between the 2 groups. There were no significant differences in hemoglobin A1c levels between the CGMS group and the SMBG group at baseline (9.1% +/- 1.1% vs 9.0% +/- 1.0%, P = .70), and both groups showed statistically significant (P < .001) and similar (P = .95) improvement in hemoglobin A1c levels after 12 weeks of study. However, the CGMS group had a significantly shorter duration of hypoglycemia (sensor glucose, < or = 60 mg/dL) at week 12 of the study (49.4 +/- 40.8 vs 81.0 +/- 61.1 minutes per event, P = .009).
Conclusion: Use of the CGMS to guide therapy adjustments in patients with insulin-treated diabetes reduces the duration of hypoglycemia compared with therapy adjustments guided by SMBG values alone.
Similar articles
-
[Usefulness of continuous glucose monitoring system (CGMS) in monitoring glycaemic profile in small children with diabetes type 1].Endokrynol Diabetol Chor Przemiany Materii Wieku Rozw. 2005;11(4):237-43. Endokrynol Diabetol Chor Przemiany Materii Wieku Rozw. 2005. PMID: 16232362 Polish.
-
Continuous glucose monitoring and the reality of metabolic control in preschool children with type 1 diabetes.Diabetes Care. 2004 Dec;27(12):2881-6. doi: 10.2337/diacare.27.12.2881. Diabetes Care. 2004. PMID: 15562201
-
Use of a real-time continuous glucose monitoring system in children and young adults on insulin pump therapy: patients' and caregivers' perception of benefit.Pediatr Diabetes. 2010 May;11(3):182-7. doi: 10.1111/j.1399-5448.2009.00549.x. Epub 2009 Dec 2. Pediatr Diabetes. 2010. PMID: 19958460
-
Vascular Glucose Sensor Symposium: Continuous Glucose Monitoring Systems (CGMS) for Hospitalized and Ambulatory Patients at Risk for Hyperglycemia, Hypoglycemia, and Glycemic Variability.J Diabetes Sci Technol. 2015 Jul;9(4):725-38. doi: 10.1177/1932296815587938. Epub 2015 Jun 15. J Diabetes Sci Technol. 2015. PMID: 26078254 Free PMC article. Review.
-
Automated Insulin Delivery-The Light at the End of the Tunnel.J Pediatr. 2017 Jul;186:17-28.e9. doi: 10.1016/j.jpeds.2017.02.055. Epub 2017 Apr 7. J Pediatr. 2017. PMID: 28396030 Review. No abstract available.
Cited by
-
A review of current evidence with continuous glucose monitoring in patients with diabetes.J Diabetes Sci Technol. 2008 Jul;2(4):718-27. doi: 10.1177/193229680800200426. J Diabetes Sci Technol. 2008. PMID: 19885251 Free PMC article.
-
Professional continuous glucose monitoring in clinical practice 2010.J Diabetes Sci Technol. 2010 Mar 1;4(2):440-56. doi: 10.1177/193229681000400226. J Diabetes Sci Technol. 2010. PMID: 20307406 Free PMC article.
-
Impact of Insulin Delivery Method on Hypoglycemia Incidence in Pediatric Type 1 Diabetes Mellitus Patients.Med Arch. 2017 Dec;71(6):391-395. doi: 10.5455/medarh.2017.71.391-395. Med Arch. 2017. PMID: 29416197 Free PMC article.
-
Efficacy of unblinded and blinded intermittently scanned continuous glucose monitoring for glycemic control in adults with type 1 diabetes.Front Endocrinol (Lausanne). 2023 Feb 22;14:1110845. doi: 10.3389/fendo.2023.1110845. eCollection 2023. Front Endocrinol (Lausanne). 2023. PMID: 36909344 Free PMC article. Clinical Trial.
-
Personal Versus Professional Continuous Glucose Monitoring: When to Use Which on Whom.Diabetes Spectr. 2019 Aug;32(3):183-193. doi: 10.2337/ds18-0093. Diabetes Spectr. 2019. PMID: 31462872 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical