Evaluation of a new combined transcutaneous measurement of PCO2/pulse oximetry oxygen saturation ear sensor in newborn patients
- PMID: 15601814
- DOI: 10.1542/peds.2004-0946
Evaluation of a new combined transcutaneous measurement of PCO2/pulse oximetry oxygen saturation ear sensor in newborn patients
Abstract
Objective: Arterial oxygen saturation (Sao(2)) and arterial carbon dioxide partial pressure (Paco(2)) are 2 of the most important respiratory parameters in the treatment of critically ill neonates. Noninvasive monitoring of these parameters is desirable for continuous estimating of the respiratory status and reducing blood loss because of repeated blood gas analyses. Transcutaneous measurement of Pco(2) (Ptcco(2)) represents a simple and noninvasive technique for continuous monitoring of ventilation. However, sensor preparation, positioning, taping, and repeated changes of the sensor location make the handling difficult and complicate its use in the neonatal care unit. Recently, a new sensor for combined assessment of pulse oximetry oxygen saturation (Spo(2)) and Ptcco(2) has been introduced (TOSCA Monitor; Linde Medical Sensors, Basel, Switzerland). The monitor combines pulse oximetry and Ptcco(2) measurement in a single ear sensor, which works at 42 degrees C to enhance blood flow in capillaries below the sensor.
Methods: In a prospective, open, nonrandomized study of 60 ill neonates, the new ear sensor for combined assessment of Spo(2) and Ptcco(2) at 42 degrees C was tested. The sensor was adapted to the ear of a neonate with a Varihesive layer (Conva Tec; Princeton, NJ). Data obtained from the ear sensor were compared with Spo(2 Finger/Heel), Sao(2), and Paco(2) obtained from arterial blood gas in 30 patients and with a capillary blood gas in an additional 30 patients using Bland Altman bias analysis. Data are presented as median (range).
Results: The postconceptional age of the patients was 38.3 weeks (range: 28 5/7-40 5/7) in the arterial group and 37.9 weeks (range: 29 6/7-41 0/7) in the capillary group. Age of the newborns studied was 3.5 days (range: 1-28) in the arterial blood sample group (n = 30) and 6 days (range: 2-28) in the capillary blood sample group (n = 30). Patient weight was 3.02 kg (range: 1.5-4.5) in the arterial group and 2.76 kg (range: 1.0-3.71) in the other group. Three patients had weights of <1500 g. Twenty-one of 60 patients were conventionally ventilated, 4 patients received high-frequency oscillation, and 35 were not ventilated. Mean difference (bias) and precision (2 SD of the mean difference) between Ptcco(2 TOSCA) and Paco(2) were -0.44 kPa (-3.21 mm Hg) and 0.82 kPa (6.02 mm Hg) and between Ptcco(2 TOSCA) and Pcapco(2) were -0.09 kPa (-0.67 mm Hg) and 1.11 kPa (8.07 mm Hg), respectively. Spo(2) assessment by the TOSCA revealed slightly higher values compared with Sao(2) (bias: -0.48%), whereas Spo(2) (Finger/Heel) values were slightly lower than Sao(2) (bias: 0.52%).
Conclusion: The TOSCA monitor with the ear sensor adapted to ears of neonates allows reliable estimation of Sao(2) and Paco(2). A potential benefit is the reduction in motion artifacts because of less head movement in newborns and that only a single cable leads form the patient to the monitor. In addition, the sensor is not removed for chest radiograph or for nursing the infant on his or her parent's lap. Long-term studies in a large population with continuous measurements are required to confirm these preliminary findings and to elucidate the benefits in detection of respiratory deterioration and the potential side effects of this sensor.
Similar articles
-
An evaluation of a new combined Spo2/PtcCO2 sensor in very low birth weight infants.Anesth Analg. 2008 Jul;107(1):125-9. doi: 10.1213/ane.0b013e3181733e47. Anesth Analg. 2008. PMID: 18635477
-
Longtime performance and reliability of two different PtcCO2 and SpO2 sensors in neonates.Paediatr Anaesth. 2008 Sep;18(9):872-7. doi: 10.1111/j.1460-9592.2008.02661.x. Paediatr Anaesth. 2008. PMID: 18768047
-
The revised digital transcutaneous PCO2/SpO2 ear sensor is a reliable noninvasive monitoring tool in patients after cardiac surgery.J Cardiothorac Vasc Anesth. 2011 Apr;25(2):243-9. doi: 10.1053/j.jvca.2010.06.021. Epub 2010 Sep 20. J Cardiothorac Vasc Anesth. 2011. PMID: 20851636
-
[Non-invasive method of arterial oxygen saturation monitoring in the neonatal intensive care unit (with a review of blood oxygen monitoring methods)].Tijdschr Kindergeneeskd. 1988 Feb;56(1):20-6. Tijdschr Kindergeneeskd. 1988. PMID: 3282343 Review. Dutch.
-
[Transcutaneous PCO2 determination in neonatology].Tijdschr Kindergeneeskd. 1991 Apr;59(2):44-50. Tijdschr Kindergeneeskd. 1991. PMID: 2053109 Review. Dutch.
Cited by
-
Guideline-oriented perioperative management of patients with bronchial asthma and chronic obstructive pulmonary disease.J Anesth. 2008;22(4):412-28. doi: 10.1007/s00540-008-0650-2. Epub 2008 Nov 15. J Anesth. 2008. PMID: 19011781 Review.
-
Monitoring in the intensive care.Crit Care Res Pract. 2012;2012:473507. doi: 10.1155/2012/473507. Epub 2012 Aug 27. Crit Care Res Pract. 2012. PMID: 22970356 Free PMC article.
-
Carbon dioxide monitoring during laparoscopic-assisted bariatric surgery in severely obese patients: transcutaneous versus end-tidal techniques.J Clin Monit Comput. 2015 Feb;29(1):183-6. doi: 10.1007/s10877-014-9587-1. Epub 2014 Jun 11. J Clin Monit Comput. 2015. PMID: 24916514
-
Use of a combined SpO₂/PtcCO₂ sensor in the delivery room.Sensors (Basel). 2012;12(8):10980-9. doi: 10.3390/s120810980. Epub 2012 Aug 8. Sensors (Basel). 2012. PMID: 23112642 Free PMC article.
-
Short term general anesthesia for retro-bulbar block in ophthalmic surgery generates no significant hypercapnia.J Clin Monit Comput. 2018 Apr;32(2):351-358. doi: 10.1007/s10877-017-0011-5. Epub 2017 Mar 11. J Clin Monit Comput. 2018. PMID: 28283852
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical
Research Materials