. 2005 Feb;26(4):376-83; discussion 325-7.
doi: 10.1093/eurheartj/ehi009. Epub 2004 Nov 29.

In vivo heating of pacemaker leads during magnetic resonance imaging

Affiliations
Free article

In vivo heating of pacemaker leads during magnetic resonance imaging

Roger Luechinger et al. Eur Heart J. 2005 Feb.
Free article

Abstract

Aims: Magnetic resonance imaging (MRI) is well established as an important diagnostic tool in medicine. However, the presence of a cardiac pacemaker is usually regarded as a contraindication for MRI due to safety reasons. In this study, heating effects at the myocardium-pacemaker lead tip interface have been investigated in a chronic animal model during MRI at 1.5 Tesla.

Methods and results: Pacemaker leads with additional thermocouple wires as temperature sensors were implanted in nine animals. Temperature increases of up to 20 degrees C were measured during MRI of the heart. Significant impedance and minor stimulation threshold changes could be seen. However, pathology and histology could not clearly demonstrate heat-induced damage.

Conclusions: MRI may produce considerable heating at the lead tip. Changes of pacing parameters due to MRI could be seen in chronic experiments. Potential risk of tissue damage cannot be excluded even though no reproducible alterations at the histological level could be found.