Metal-on-metal bearings and hypersensitivity in patients with artificial hip joints. A clinical and histomorphological study
- PMID: 15637030
- DOI: 10.2106/JBJS.A.02039pp
Metal-on-metal bearings and hypersensitivity in patients with artificial hip joints. A clinical and histomorphological study
Abstract
Background: Some patients who have a total hip replacement with a second-generation metal-on-metal articulation have persistent or early recurrence of preoperative symptoms. Characteristic histological changes in the periprosthetic tissues suggested the development of an immunological response. Therefore, in order to determine the relevance of these symptoms, we performed a study of the clinical data and periprosthetic tissues associated with endoprostheses with a metal-on metal articulation that had been retrieved at revision.
Methods: Periprosthetic tissues as well as the clinical data on the patients were obtained from the first nineteen consecutive revisions performed at the treating hospitals. At the time of the revision, fourteen patients had the metal-on-metal articulation exchanged for either an alumina-ceramic or a metal-on-polyethylene articulation. Five patients received another second-generation metal-on-metal total joint replacement. Five-micrometer sections were prepared from the tissue samples, were stained with routine and immunohistochemical methods, and were examined histologically. Histological specimens from three groups of patients, two of which were treated with non-metal-on-metal implants, served as controls.
Results: The majority of patients had persistence of their preoperative pain or early recurrence of the pain after the original total hip replacement, and often a pronounced hip joint effusion had developed after the original replacement. Radiographic follow-up showed the development of radiolucent lines in five hips and of osteolysis in another seven hips. At the revision surgery, both the cup and the stem were found to be well fixed in nine patients. The characteristic histological features were diffuse and perivascular infiltrates of T and B lymphocytes and plasma cells, high endothelial venules, massive fibrin exudation, accumulation of macrophages with droplike inclusions, and infiltrates of eosinophilic granulocytes and necrosis. Only a few metal particles were detected. Immunohistochemical analysis demonstrated that the cellular reaction was still active. The patients who received another second-generation metal-on-metal articulation at the time of the revision had no decrease in symptoms. In the control group of tissues obtained at revisions of endoprostheses without cobalt, chromium, or nickel articulations, there were no similar signs of immune reactions.
Conclusions: These histological findings support the possibility of a lymphocyte-dominated immunological response. Although the prevalence of this reaction is low, the persistence or early reappearance of symptoms, including a marked joint effusion and the development of osteolysis, after primary implantation may suggest the possibility of such a reaction.
Similar articles
-
An unusual lymphocytic perivascular infiltration in tissues around contemporary metal-on-metal joint replacements.J Bone Joint Surg Am. 2005 Jan;87(1):18-27. doi: 10.2106/JBJS.C.00949. J Bone Joint Surg Am. 2005. PMID: 15634811
-
Immunologic adverse reaction associated with low-carbide metal-on-metal bearings in total hip arthroplasty.Clin Orthop Relat Res. 2010 Aug;468(8):2135-42. doi: 10.1007/s11999-009-1187-x. Epub 2009 Dec 18. Clin Orthop Relat Res. 2010. PMID: 20020335 Free PMC article.
-
Early osteolysis following second-generation metal-on-metal hip replacement.J Bone Joint Surg Am. 2005 Jul;87(7):1515-21. doi: 10.2106/JBJS.D.02641. J Bone Joint Surg Am. 2005. PMID: 15995119
-
Metallosis after contemporary metal-on-metal total hip arthroplasty. Five to nine-year follow-up.J Bone Joint Surg Am. 2006 Jun;88(6):1183-91. doi: 10.2106/JBJS.D.02916. J Bone Joint Surg Am. 2006. PMID: 16757749
-
Understanding outcomes and toxicological aspects of second generation metal-on-metal hip implants: a state-of-the-art review.Crit Rev Toxicol. 2018 Nov;48(10):853-901. doi: 10.1080/10408444.2018.1563048. Epub 2019 Mar 26. Crit Rev Toxicol. 2018. PMID: 30912993 Review.
Cited by
-
Cementation and interface analysis of early failure cases after hip-resurfacing arthroplasty.Int Orthop. 2012 Jul;36(7):1333-40. doi: 10.1007/s00264-011-1464-7. Epub 2012 Jan 11. Int Orthop. 2012. PMID: 22234704 Free PMC article.
-
Adverse Local Tissue Reaction Secondary to Corrosion at Multiple Junctions in a Modular, Segmental, Distal Femoral Replacement.Arthroplast Today. 2023 Nov 9;24:101256. doi: 10.1016/j.artd.2023.101256. eCollection 2023 Dec. Arthroplast Today. 2023. PMID: 38023655 Free PMC article.
-
Biomaterial hypersensitivity: is it real? Supportive evidence and approach considerations for metal allergic patients following total knee arthroplasty.Biomed Res Int. 2015;2015:137287. doi: 10.1155/2015/137287. Epub 2015 Mar 25. Biomed Res Int. 2015. PMID: 25883940 Free PMC article. Review.
-
Effect of femoral head size on risk of revision for dislocation after total hip arthroplasty: a population-based analysis of 42,379 primary procedures from the Finnish Arthroplasty Register.Acta Orthop. 2013 Aug;84(4):342-7. doi: 10.3109/17453674.2013.810518. Epub 2013 Jun 25. Acta Orthop. 2013. PMID: 23799348 Free PMC article.
-
The effect of contact load on CoCrMo wear and the formation and retention of tribofilms.Wear. 2015 May-Jun;332-333:643-649. doi: 10.1016/j.wear.2015.02.013. Wear. 2015. PMID: 26085697 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical