Direct real-time measurement of in vivo forces in the lumbar spine
- PMID: 15653089
- DOI: 10.1016/j.spinee.2004.06.017
Direct real-time measurement of in vivo forces in the lumbar spine
Abstract
Background context: Accurate knowledge of the mechanical loads in the lumbar spine is critical to understanding the causes of degenerative disc disease and to developing suitable treatment options and functional disc replacements. To date, only indirect methods have been used to measure the forces developed in the spine in vivo. These methods are fraught with error, and results have never been validated using direct experimental measurements.
Purpose: The major aims of this study were to develop a methodology to directly measure, in real time, the in vivo loading in the lumbar spine, to determine if the forces developed in the lumbar spine are dependent on activity and/or posture and to assess the baboon as an animal model for human lumbar spine research based on in vivo mechanical loading.
Study design: Real-time telemetered data were collected from sensor-imbedded implants that were placed in the interbody space of the lumbar spines of two baboons.
Methods: An interbody spinal implant was designed and instrumented with strain gauges to be used as a load cell. The implant was placed anteriorly in the lumbar spine of the baboon. Strain data were collected in vivo during normal activities and transmitted by means of a telemetry system to a receiver. The forces transmitted through the implant were calculated from the measured strain based on precalibration of the load cell. Measured forces were correlated to videotaped activities to elucidate trends in force level as a function of activity and posture over a 6-week period. The procedure was repeated in a second baboon, and data were recorded for similar activities.
Results: Implants measured in vivo forces developed in the lumbar spine with less than 10% error. Loads in the lumbar spine are dependent on activity and posture. The maximum loads developed in the lumbar spine during normal (baboon) activities exceeded four times body weight and were recorded while animals were sitting flexed. Force data indicate similar trends between the human lumbar spine and the baboon lumbar spine.
Conclusions: It is possible to monitor the real-time forces present in the lumbar spine. Force data correlate well to trends previously reported for in vivo pressure data. Results also indicate that the baboon may be an appropriate animal model for study of the human lumbar spine.
Similar articles
-
Real-time in vivo loading in the lumbar spine: part 1. Interbody implant: load cell design and preliminary results.Spine (Phila Pa 1976). 2000 Oct 15;25(20):2595-600. doi: 10.1097/00007632-200010150-00009. Spine (Phila Pa 1976). 2000. PMID: 11034643
-
A biomechanical assessment of disc pressures in the lumbosacral spine in response to external unloading forces.Spine J. 2005 Sep-Oct;5(5):548-53. doi: 10.1016/j.spinee.2005.03.012. Spine J. 2005. PMID: 16153584
-
Two in vivo surgical approaches for lumbar corpectomy using allograft and a metallic implant: a controlled clinical and biomechanical study.Spine J. 2006 Nov-Dec;6(6):648-58. doi: 10.1016/j.spinee.2006.04.028. Epub 2006 Oct 11. Spine J. 2006. PMID: 17088195
-
Estimation of loads on human lumbar spine: A review of in vivo and computational model studies.J Biomech. 2016 Apr 11;49(6):833-845. doi: 10.1016/j.jbiomech.2015.12.038. Epub 2015 Dec 31. J Biomech. 2016. PMID: 26873281 Review.
-
Durability/wear testing of heart valve substitutes.J Heart Valve Dis. 1998 Mar;7(2):151-7. J Heart Valve Dis. 1998. PMID: 9587854 Review.
Cited by
-
Sitting Posture, Sagittal Spinal Curvatures and Back Pain in 8 to 12-Year-Old Children from the Region of Murcia (Spain): ISQUIOS Programme.Int J Environ Res Public Health. 2020 Apr 9;17(7):2578. doi: 10.3390/ijerph17072578. Int J Environ Res Public Health. 2020. PMID: 32283720 Free PMC article.
-
THREE-DIMENSIONAL MULTI-SEGMENTED SPINE JOINT REACTION FORCES DURING COMMON WORKPLACE PHYSICAL DEMANDS/ACTIVITIES OF DAILY LIVING.Biomed Eng (Singapore). 2017;29(4):1750025. doi: 10.4015/S1016237217500259. Epub 2017 Aug 14. Biomed Eng (Singapore). 2017. PMID: 30820137 Free PMC article.
-
The application of impantable sensors in the musculoskeletal system: a review.Front Bioeng Biotechnol. 2024 Jan 24;12:1270237. doi: 10.3389/fbioe.2024.1270237. eCollection 2024. Front Bioeng Biotechnol. 2024. PMID: 38328442 Free PMC article. Review.
-
Biomechanical and in vivo evaluation of experimental closure devices of the annulus fibrosus designed for a goat nucleus replacement model.Eur Spine J. 2010 Aug;19(8):1347-55. doi: 10.1007/s00586-010-1384-z. Epub 2010 Apr 17. Eur Spine J. 2010. PMID: 20401620 Free PMC article.
-
Comparison of lumbo-pelvic kinematics during trunk forward bending and backward return between patients with acute low back pain and asymptomatic controls.Clin Biomech (Bristol). 2017 Jan;41:66-71. doi: 10.1016/j.clinbiomech.2016.12.005. Epub 2016 Dec 10. Clin Biomech (Bristol). 2017. PMID: 27992778 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources