[Digital mammography with high-resolution storage plates (CR) versus full-field digital mammography (CCD) (DR) for microcalcifications and focal lesions -- a retrospective clinical histologic analysis (n = 102)]
- PMID: 15657822
- DOI: 10.1055/s-2004-813722
[Digital mammography with high-resolution storage plates (CR) versus full-field digital mammography (CCD) (DR) for microcalcifications and focal lesions -- a retrospective clinical histologic analysis (n = 102)]
Abstract
Purpose: To determine the diagnostic accuracy of microcalcifications and focal lesions in a retrospective clinical-histological study using high-resolution digital phosphor storage plates (hard copy) and full-field digital mammography (hard copy).
Materials and methods: From May 2003 to September 2003, 102 patients underwent digital storage plate mammography (CR), using a mammography unit (Mammomat 3000 N, Siemens) in combination with a high resolution (9 lp/mm) digital storage phosphor plate system (pixel size 50 microm) (Fuji/Siemens). After diagnosis and preoperative wire localization, full-field digital mammography (CCD) (DR) was performed with the same exposure parameters. The full-field digital mammography used a CCD-detector (SenoScan) (Fisher Imaging) with a resolution of 10 Ip/mm and a pixel size of 50 microm. Five investigators determined the diagnosis (BI-RADS I - V) retrospectively after the operation from randomly distributed mediolateral views (hard copy reading). These results were correlated with the final histology.
Results: The diagnostic accuracy of digital storage plate mammography (CR) and full-field digital mammography (CCD) (DR) was 73 % and 71 % for all findings (n = 102), 73 % and 71 % for microcalcifications (n = 51), and 72 % and 70 % for focal lesions (n = 51). The overall results showed no difference.
Conclusion: Our findings indicate the equivalence of high-resolution digital phosphor storage plate mammography (CR) and full-field digital mammography (CCD) (DR).
Similar articles
-
[Film-screen mammography versus digital storage plate mammography: hard copy and monitor display of microcalcifications and focal findings--a retrospective clinical and histologic analysis].Rofo. 2003 Sep;175(9):1220-4. doi: 10.1055/s-2003-41928. Rofo. 2003. PMID: 12964077 German.
-
[DR (a-Se) versus CR (DLR)--is an improvement of the accuracy possible? A retrospective histologic analysis (n = 100)].Rontgenpraxis. 2007;56(4):129-35. doi: 10.1016/j.rontge.2006.10.001. Rontgenpraxis. 2007. PMID: 17390952 German.
-
[Comparison of dignity determination of mammographic microcalcification with two systems for digital full-field mammography with different detector resolution: a retrospective clinical study].Radiologe. 2011 Feb;51(2):126-9. doi: 10.1007/s00117-010-2078-6. Radiologe. 2011. PMID: 21069511 German.
-
[Digital magnification mammography. A new technique for improved visualization of microcalcifications in breast cancer diagnosis].Radiologe. 1993 May;33(5):260-6. Radiologe. 1993. PMID: 8516436 Review. German.
-
[Full-field digital mammography].Rofo. 2000 Dec;172(12):957-64. doi: 10.1055/s-2000-9210. Rofo. 2000. PMID: 11199437 Review. German.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical