Imaging at 3.4 Thz with a quantum-cascade laser
- PMID: 15662893
- DOI: 10.1364/ao.44.000121
Imaging at 3.4 Thz with a quantum-cascade laser
Abstract
We have assembled a single-frequency imaging system at 3.4 THz with a quantum-cascade laser. Images of electronic and biological applications are demonstrated. We operate the laser with a peak output power of 2.5 mW at a 7% duty cycle and a 22 K operating temperature. The minimum spot size is 340 microm. The signal is detected with a single-element deuterated triglycine sulfate detector, and images are captured by scanning of the sample.