Glaucomatous visual field progression with frequency-doubling technology and standard automated perimetry in a longitudinal prospective study
- PMID: 15671281
- DOI: 10.1167/iovs.04-0973
Glaucomatous visual field progression with frequency-doubling technology and standard automated perimetry in a longitudinal prospective study
Abstract
Purpose: To compare frequency-doubling technology (FDT) perimetry with standard automated perimetry (SAP) for detecting glaucomatous visual field progression in a longitudinal prospective study.
Methods: One eye of patients with open-angle glaucoma was tested every 6 months with both FDT and SAP. A minimum of 6 examinations with each perimetric technique was required for inclusion. Visual field progression was determined by two methods: glaucoma change probability (GCP) analysis and linear regression analysis (LRA). For GCP, several criteria for progression were used. The number of locations required to classify progression with FDT compared with SAP, respectively, was 1:2 (least conservative), 1:3, 2:3, 2:4, 2:6, 2:7, 3:6, 3:7, and 3:10 (most conservative). The number of consecutive examinations required to confirm progression was 2-of-3, 2-of-2, and 3-of-3. For LRA, the progression criterion was any significant decline in mean threshold sensitivity over time in each of the following three visual field subdivisions: (1) all test locations, (2) locations in the central 10 degrees and the superior and inferior hemifields, and (3) locations in each quadrant. Using these criteria, the proportion of patients classified as showing progression with each perimetric technique was calculated and, in the case of progression with both, the differences in time to progression were determined.
Results: Sixty-five patients were followed for a median of 3.5 years (median number of examinations, 9). For the least conservative GCP criterion, 32 (49%) patients were found to have progressing visual fields with FDT and 32 (49%) patients with SAP. Only 16 (25%) patients showed progression with both methods, and in most of those patients, FDT identified progression before SAP (median, 12 months earlier). The majority of GCP progression criteria (15/27), classified more patients as showing progression with FDT than with SAP. Contrary to this, more patients showed progression with SAP than FDT, when analysed with LRA; e.g., using quadrant LRA 20 (31%) patients showed progression with FDT, 23 (35%) with SAP, and only 10 (15%) with both.
Conclusions: FDT perimetry detected glaucomatous visual field progression. However, the proportion of patients who showed progression with both FDT and SAP was small, possibly indicating that the two techniques identify different subgroups of patients. Using GCP, more patients showed progression with FDT than with SAP, yet the opposite occurred using LRA. As there is no independent qualifier of progression, FDT and SAP progression rates vary depending on the method of analysis and the criterion used.
Similar articles
-
Usefulness of frequency-doubling technology for perimetrically normal eyes of open-angle glaucoma patients with unilateral field loss.Ophthalmology. 2010 Aug;117(8):1530-7, 1537.e1-2. doi: 10.1016/j.ophtha.2009.12.034. Epub 2010 May 13. Ophthalmology. 2010. PMID: 20466428
-
Can frequency-doubling technology and short-wavelength automated perimetries detect visual field defects before standard automated perimetry in patients with preperimetric glaucoma?J Glaucoma. 2007 Jun-Jul;16(4):372-83. doi: 10.1097/IJG.0b013e31803bbb17. J Glaucoma. 2007. PMID: 17571000
-
Relationship of SITA and full-threshold standard perimetry to frequency-doubling technology perimetry in glaucoma.Invest Ophthalmol Vis Sci. 2005 Jul;46(7):2433-9. doi: 10.1167/iovs.04-1108. Invest Ophthalmol Vis Sci. 2005. PMID: 15980232
-
Short-wavelength automated perimetry and frequency-doubling technology perimetry in glaucoma.Prog Brain Res. 2008;173:101-24. doi: 10.1016/S0079-6123(08)01108-4. Prog Brain Res. 2008. PMID: 18929104 Review.
-
[Diagnostic approaches for early detection of glaucoma progression].Klin Monbl Augenheilkd. 2006 Mar;223(3):194-216. doi: 10.1055/s-2005-858734. Klin Monbl Augenheilkd. 2006. PMID: 16552653 Review. German.
Cited by
-
Predicting progression of glaucoma from rates of frequency doubling technology perimetry change.Ophthalmology. 2014 Feb;121(2):498-507. doi: 10.1016/j.ophtha.2013.09.016. Epub 2013 Nov 26. Ophthalmology. 2014. PMID: 24289917 Free PMC article.
-
Comparison of Standard Automated Perimetry, Short-Wavelength Automated Perimetry, and Frequency-Doubling Technology Perimetry to Monitor Glaucoma Progression.Medicine (Baltimore). 2016 Feb;95(7):e2618. doi: 10.1097/MD.0000000000002618. Medicine (Baltimore). 2016. PMID: 26886602 Free PMC article.
-
Practical recommendations for measuring rates of visual field change in glaucoma.Br J Ophthalmol. 2008 Apr;92(4):569-73. doi: 10.1136/bjo.2007.135012. Epub 2008 Jan 22. Br J Ophthalmol. 2008. PMID: 18211935 Free PMC article. Review.
-
Functional assessment of glaucoma: Uncovering progression.Surv Ophthalmol. 2020 Nov-Dec;65(6):639-661. doi: 10.1016/j.survophthal.2020.04.004. Epub 2020 Apr 26. Surv Ophthalmol. 2020. PMID: 32348798 Free PMC article. Review.
-
Comparison of matrix frequency-doubling technology perimetry and standard automated perimetry in monitoring the development of visual field defects for glaucoma suspect eyes.PLoS One. 2017 May 18;12(5):e0178079. doi: 10.1371/journal.pone.0178079. eCollection 2017. PLoS One. 2017. PMID: 28542536 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical
Research Materials
Miscellaneous