Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Review
. 2005 Feb;50(2):202-34; discussion 232-4.

Using ventilator graphics to identify patient-ventilator asynchrony

Affiliations
  • PMID: 15691392
Review

Using ventilator graphics to identify patient-ventilator asynchrony

Jon O Nilsestuen et al. Respir Care. 2005 Feb.

Abstract

Patient-ventilator interaction can be described as the relationship between 2 respiratory pumps: (1) the patient's pulmonary system, which is controlled by the neuromuscular system and influenced by the mechanical characteristics of the lungs and thorax, and (2) the ventilator, which is controlled by the ventilator settings and the function of the flow valve. When the 2 pumps function in synchrony, every phase of the breath is perfectly matched. Anything that upsets the harmony between the 2 pumps results in asynchrony and causes patient discomfort and unnecessarily increases work of breathing. This article discusses asynchrony relative to the 4 phases of a breath and illustrates how asynchrony can be identified with the 3 standard ventilator waveforms: pressure, flow, and volume. The 4 phases of a breath are: (1) The trigger mechanism (ie, initiation of the inspiration), which is influenced by the trigger-sensitivity setting, patient effort, and valve responsiveness. (2) The inspiratory-flow phase. During both volume-controlled and pressure-controlled ventilation the patient's flow demand should be carefully evaluated, using the pressure and flow waveforms. (3) Breath termination (ie, the end of the inspiration). Ideally, the ventilator terminates inspiratory flow in synchrony with the patient's neural timing, but frequently the ventilator terminates inspiration either early or late, relative to the patient's neural timing. During volume-controlled ventilation we can adjust variables that affect inspiratory time (eg, peak flow, tidal volume). During pressure-controlled or pressure-support ventilation we can adjust variables that affect when the inspiration terminates (eg, inspiratory time, expiratory sensitivity). (4) Expiratory phase. Patients with obstructive lung disease are particularly prone to developing intrinsic positive end-expiratory pressure (auto-PEEP) and therefore have difficulty triggering the ventilator. Bedside evaluation for the presence of auto-PEEP should be routinely performed and corrective adjustments made when appropriate.

PubMed Disclaimer

Similar articles

Cited by

MeSH terms

LinkOut - more resources