Using ventilator graphics to identify patient-ventilator asynchrony
- PMID: 15691392
Using ventilator graphics to identify patient-ventilator asynchrony
Abstract
Patient-ventilator interaction can be described as the relationship between 2 respiratory pumps: (1) the patient's pulmonary system, which is controlled by the neuromuscular system and influenced by the mechanical characteristics of the lungs and thorax, and (2) the ventilator, which is controlled by the ventilator settings and the function of the flow valve. When the 2 pumps function in synchrony, every phase of the breath is perfectly matched. Anything that upsets the harmony between the 2 pumps results in asynchrony and causes patient discomfort and unnecessarily increases work of breathing. This article discusses asynchrony relative to the 4 phases of a breath and illustrates how asynchrony can be identified with the 3 standard ventilator waveforms: pressure, flow, and volume. The 4 phases of a breath are: (1) The trigger mechanism (ie, initiation of the inspiration), which is influenced by the trigger-sensitivity setting, patient effort, and valve responsiveness. (2) The inspiratory-flow phase. During both volume-controlled and pressure-controlled ventilation the patient's flow demand should be carefully evaluated, using the pressure and flow waveforms. (3) Breath termination (ie, the end of the inspiration). Ideally, the ventilator terminates inspiratory flow in synchrony with the patient's neural timing, but frequently the ventilator terminates inspiration either early or late, relative to the patient's neural timing. During volume-controlled ventilation we can adjust variables that affect inspiratory time (eg, peak flow, tidal volume). During pressure-controlled or pressure-support ventilation we can adjust variables that affect when the inspiration terminates (eg, inspiratory time, expiratory sensitivity). (4) Expiratory phase. Patients with obstructive lung disease are particularly prone to developing intrinsic positive end-expiratory pressure (auto-PEEP) and therefore have difficulty triggering the ventilator. Bedside evaluation for the presence of auto-PEEP should be routinely performed and corrective adjustments made when appropriate.
Similar articles
-
Ventilator waveforms and the physiology of pressure support ventilation.Respir Care. 2005 Feb;50(2):166-86; discussion 183-6. Respir Care. 2005. PMID: 15691390 Review.
-
Ventilator graphics and respiratory mechanics in the patient with obstructive lung disease.Respir Care. 2005 Feb;50(2):246-61; discussion 259-61. Respir Care. 2005. PMID: 15691394 Review.
-
[Patient-ventilator interaction].Tuberk Toraks. 2009;57(4):453-65. Tuberk Toraks. 2009. PMID: 20037864 Review. Turkish.
-
Cycling of the mechanical ventilator breath.Respir Care. 2011 Jan;56(1):52-60. doi: 10.4187/respcare.01088. Respir Care. 2011. PMID: 21235838
-
Monitoring of patient-ventilator interaction at the bedside.Respir Care. 2011 Jan;56(1):61-72. doi: 10.4187/respcare.01077. Respir Care. 2011. PMID: 21235839
Cited by
-
Monitoring Patient/Ventilator Interactions: Manufacturer's Perspective.Open Respir Med J. 2009 Mar 12;3:17-26. doi: 10.2174/1874306400903010017. Open Respir Med J. 2009. PMID: 19452035 Free PMC article.
-
Reverse triggering ? a novel or previously missed phenomenon?Ann Intensive Care. 2024 May 22;14(1):78. doi: 10.1186/s13613-024-01303-4. Ann Intensive Care. 2024. PMID: 38776032 Free PMC article.
-
The Ability of Critical Care Physicians to Identify Patient-Ventilator Asynchrony Using Waveform Analysis: A National Survey.Respir Care. 2024 Jan 24;69(2):176-183. doi: 10.4187/respcare.11360. Respir Care. 2024. PMID: 38267232 Free PMC article.
-
The importance of characterising dynamic response and inertia in potential rapidly manufactured ventilator systems.Anaesthesia. 2020 Dec;75(12):1690-1691. doi: 10.1111/anae.15190. Epub 2020 Jul 4. Anaesthesia. 2020. PMID: 32621288 Free PMC article. No abstract available.
-
Not all types of asynchrony are created equal.Intensive Care Med. 2013 Feb;39(2):338. doi: 10.1007/s00134-012-2737-3. Epub 2012 Oct 26. Intensive Care Med. 2013. PMID: 23100009 No abstract available.
Publication types
MeSH terms
LinkOut - more resources
Other Literature Sources