Scanning laser polarimetry retinal nerve fiber layer thickness measurements after LASIK
- PMID: 15691551
- DOI: 10.1016/j.ophtha.2004.08.019
Scanning laser polarimetry retinal nerve fiber layer thickness measurements after LASIK
Abstract
Purpose: To compare retinal nerve fiber layer (RNFL) thickness measurements before and after LASIK.
Design: Cohort study.
Participants: Twenty participants undergoing LASIK and 14 normal controls.
Methods: Retinal nerve fiber layer thickness was measured before LASIK and approximately 3 months after surgery in one eye each of 20 patients using a scanning laser polarimeter (GDx Nerve Fiber Analyzer) with fixed corneal compensation (FCC), one with variable corneal compensation (GDx VCC), and optical coherence tomography (OCT). Fourteen normal controls also were tested at baseline and approximately 3 months later.
Main outcome measures: Retinal nerve fiber layer thicknesses measured with the GDx FCC, GDx VCC, and OCT.
Results: At baseline, mean (95% confidence interval [CI]) RNFL thicknesses for the GDx FCC, GDx VCC, and OCT were 78.1 microm (72.2-83.9), 54.3 microm (52.7-56.0), and 96.8 microm (93.2-100.5), respectively. In both LASIK and control groups, there were no significant changes between baseline and follow-up examinations in GDx VCC and OCT RNFL thickness measurements globally or in the superior and inferior quadrants (mean change, <5 microm for each instrument). In the control group, there also was no significant change in GDx FCC measurements between baseline and follow-up. In LASIK patients, significant reductions were observed in GDx FCC RNFL measurements. Average absolute values of the mean (95% CI) change in thickness were 12.4 microm (7.7-17.2), 15.3 microm (9.6-20.9), and 12.9 microm (7.6-18.1) for GDx FCC RNFL measurements superiorly, inferiorly, and globally, respectively (all Ps < or = 0.001).
Conclusions: LASIK does not seem to change RNFL thickness. Reduction in GDx FCC RNFL thickness measurements after LASIK is a measurement artifact and is most likely due to erroneous compensation for corneal birefringence. With scanning laser polarimetry, it is mandatory to compensate individually for change in corneal birefringence after LASIK to ensure accurate RNFL assessment.
Similar articles
-
Evaluation of the retinal nerve fiber layer measurements, after photorefractive keratectomy and laser in situ keratomileusis, using scanning laser polarimetry (GDX VCC).Graefes Arch Clin Exp Ophthalmol. 2010 May;248(5):731-6. doi: 10.1007/s00417-009-1273-5. Epub 2010 Jan 14. Graefes Arch Clin Exp Ophthalmol. 2010. PMID: 20072786 Clinical Trial.
-
Retinal nerve fiber layer measurements do not change after LASIK for high myopia as measured by scanning laser polarimetry with custom compensation.Ophthalmology. 2005 Jan;112(1):92-7. doi: 10.1016/j.ophtha.2004.07.027. Ophthalmology. 2005. PMID: 15629826
-
Scanning laser polarimetry and retinal thickness analysis before and after laser in situ keratomileusis.Eur J Ophthalmol. 2005 Jul-Aug;15(4):434-40. Eur J Ophthalmol. 2005. PMID: 16001373
-
Scanning laser polarimetry with variable corneal compensation and optical coherence tomography in normal and glaucomatous eyes.Am J Ophthalmol. 2003 Apr;135(4):521-9. doi: 10.1016/s0002-9394(02)02077-9. Am J Ophthalmol. 2003. PMID: 12654370
-
Imaging of the retinal nerve fibre layer for glaucoma.Br J Ophthalmol. 2009 Feb;93(2):139-43. doi: 10.1136/bjo.2008.145540. Epub 2008 Nov 21. Br J Ophthalmol. 2009. PMID: 19028735 Free PMC article. Review.
Cited by
-
Transient effect of suction on the retinal neurovasculature in myopic patients after small-incision lenticule extraction.J Cataract Refract Surg. 2020 Feb;46(2):250-259. doi: 10.1016/j.jcrs.2019.09.003. J Cataract Refract Surg. 2020. PMID: 32126039 Free PMC article.
-
Influence of corneal power on circumpapillary retinal nerve fiber layer and optic nerve head measurements by spectral-domain optical coherence tomography.Int J Ophthalmol. 2017 Sep 18;10(9):1385-1391. doi: 10.18240/ijo.2017.09.09. eCollection 2017. Int J Ophthalmol. 2017. PMID: 28944197 Free PMC article.
-
Changes in retinal nerve fiber layer thickness following mechanical microkeratome-assisted versus femtosecond laser-assisted LASIK.Clin Ophthalmol. 2013;7:1919-22. doi: 10.2147/OPTH.S51774. Epub 2013 Sep 25. Clin Ophthalmol. 2013. PMID: 24109171 Free PMC article.
-
Evaluation of the retinal nerve fiber layer measurements, after photorefractive keratectomy and laser in situ keratomileusis, using scanning laser polarimetry (GDX VCC).Graefes Arch Clin Exp Ophthalmol. 2010 May;248(5):731-6. doi: 10.1007/s00417-009-1273-5. Epub 2010 Jan 14. Graefes Arch Clin Exp Ophthalmol. 2010. PMID: 20072786 Clinical Trial.
-
Assessing the relationship between central corneal thickness and retinal nerve fiber layer thickness in healthy subjects.Am J Ophthalmol. 2008 Oct;146(4):561-6. doi: 10.1016/j.ajo.2008.05.038. Epub 2008 Jul 26. Am J Ophthalmol. 2008. PMID: 18657796 Free PMC article. Clinical Trial.
Publication types
MeSH terms
Grants and funding
LinkOut - more resources
Full Text Sources
Medical