Problems created in attenuation-corrected SPECT images by artifacts in attenuation maps: a simulation study
- PMID: 15695795
Problems created in attenuation-corrected SPECT images by artifacts in attenuation maps: a simulation study
Abstract
The importance of accurate attenuation correction, especially for imaging of the thorax region, is widely acknowledged. Appropriate compensation methods have been developed and introduced into clinical practice. Most of these methods use attenuation maps obtained using various transmission scanning systems. However, when maps are inaccurate, the correction procedure may introduce artifacts into the final images that can be difficult to identify and might inadvertently alter diagnosis and study outcome. As a result, attenuation correction is often avoided in clinical practice. Our objective was to examine issues related to the quality of attenuation maps and the effects that map artifacts may have on attenuation-corrected emission images.
Methods: The topics that are investigated include the problem of low transmission counts, cross-talk contributions from the emission isotope, truncation of the transmission data, and methods of map reconstruction and segmentation. Examples of patient studies displaying specific problems guided our investigations, but, because truth in these studies is seldom known, analytic and Monte Carlo-simulated data were used in the analysis. Attenuation maps and final emission images were visually checked for artifacts and for the presence of perfusion defects. In addition, quantitative evaluation of map uniformity, defect visibility, and size variation was performed.
Results: The statistical paired-sample t test showed significant (P < 0.05) improvement of relative SD for attenuation maps reconstructed with iterative methods as compared with filtered backprojection and for maps created with higher photon fluxes. When maps with artifacts were used to correct emission data, an increase in myocardial infarct size and creation of false heart defects were observed.
Conclusion: Our study strongly recommends that at least a visual inspection of the quality of attenuation maps be performed before their use in compensation procedures. To improve image quality, remove artifacts, and increase diagnostic confidence, attenuation maps used in the correction procedure must be accurate and free of artifacts.
Similar articles
-
A method to remove artifacts in attenuation-corrected myocardial perfusion SPECT Introduced by misalignment between emission scan and CT-derived attenuation maps.J Nucl Med. 2004 Oct;45(10):1619-25. J Nucl Med. 2004. PMID: 15471824 Clinical Trial.
-
Monte Carlo-based down-scatter correction of SPECT attenuation maps.Eur J Nucl Med Mol Imaging. 2004 Aug;31(8):1173-81. doi: 10.1007/s00259-004-1507-z. Epub 2004 Mar 18. Eur J Nucl Med Mol Imaging. 2004. PMID: 15034678
-
Correction of photon attenuation and collimator response for a body-contouring SPECT/CT imaging system.J Nucl Med. 2005 May;46(5):868-77. J Nucl Med. 2005. PMID: 15872362 Clinical Trial.
-
Attenuation correction: eternal dilemma or real improvement?Q J Nucl Med Mol Imaging. 2005 Mar;49(1):30-42. Q J Nucl Med Mol Imaging. 2005. PMID: 15724134 Review.
-
Determination of the attenuation map in emission tomography.J Nucl Med. 2003 Feb;44(2):291-315. J Nucl Med. 2003. PMID: 12571222 Review.
Cited by
-
Brain single-photon emission CT physics principles.AJNR Am J Neuroradiol. 2008 Aug;29(7):1247-56. doi: 10.3174/ajnr.A1175. Epub 2008 Jun 26. AJNR Am J Neuroradiol. 2008. PMID: 18583408 Free PMC article. Review.
-
Influence of respiratory motion correction on quantification of myocardial perfusion SPECT.J Nucl Cardiol. 2015 Oct;22(5):1019-30. doi: 10.1007/s12350-014-0031-1. Epub 2014 Dec 17. J Nucl Cardiol. 2015. PMID: 25515421
-
Reconstruction from uniformly attenuated SPECT projection data using the DBH method.IEEE Trans Med Imaging. 2009 Jan;28(1):17-29. doi: 10.1109/TMI.2008.923974. IEEE Trans Med Imaging. 2009. PMID: 19116185 Free PMC article.
-
Nonlinear dual reconstruction of SPECT activity and attenuation images.PLoS One. 2014 Sep 16;9(9):e106951. doi: 10.1371/journal.pone.0106951. eCollection 2014. PLoS One. 2014. PMID: 25225796 Free PMC article.
-
Phantom evaluation of a cardiac SPECT/VCT system that uses a common set of solid-state detectors for both emission and transmission scans.J Nucl Cardiol. 2010 Jun;17(3):459-69. doi: 10.1007/s12350-010-9204-8. Epub 2010 Feb 19. J Nucl Cardiol. 2010. PMID: 20169476 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Research Materials