Comparative Study
. 2005 Feb;123(2):179-85.
doi: 10.1001/archopht.123.2.179.

Noncontact goniometry with optical coherence tomography

Affiliations
Comparative Study

Noncontact goniometry with optical coherence tomography

Christopher Wirbelauer et al. Arch Ophthalmol. 2005 Feb.

Abstract

Objective: To assess the value of noncontact goniometry with optical coherence tomography (OCT) compared with current clinical parameters in the evaluation of the anterior chamber angle (ACA).

Design: Prospective observational study of 138 eyes of 109 patients.

Methods: The ACA parameters and angle-opening distance (AOD) were measured with slitlamp-adapted OCT goniometry. The iris and scleral thickness and the iris convexity were assessed with OCT. Both ACA and AOD were compared with the clinical parameters of gonioscopy grade, limbal anterior chamber depth (ACD), ultrasonographic central ACD, and lens-axial length (LAL) ratio.

Results: Noncontact goniometry with OCT revealed mean +/- SD values of 28 degrees +/- 16 degrees for the ACA and 381 +/- 234 mum for the AOD. The mean +/- SD iris thickness was 369 +/- 84 mum, and the scleral thickness at the scleral spur was 943 +/- 148 mum. There was a significant correlation (P<.001) with the clinical parameters of gonioscopic grading, limbal ACD, ultrasonographic central ACD, and LAL ratio. The sensitivity and specificity of OCT goniometry to detect an occludable angle were 86% and 95% for ACA and 85% and 90% for AOD, respectively.

Conclusions: Noncontact goniometry with OCT was helpful in evaluating the anterior chamber structures and as a screening modality. Goniometry with OCT could improve the noninvasive clinical assessment and treatment of patients with glaucoma.